![Tektronix TDS7054 Service Manual Download Page 140](http://html1.mh-extra.com/html/tektronix/tds7054/tds7054_service-manual_1078193140.webp)
Performance Tests
4- 80
TDS7104 & TDS7054 Service Manual
H
Record the setting of the DC generator.
H
Adjust the output of the DC calibration generator until it precisely
overlaps the base (lower) level of the stored probe compensation
signal. (This value will be near zero volts).
H
Record the setting of the DC generator.
f.
Press
Close
to remove the menus from the display. See Figure 4--38.
Figure 4- 38: Measurement of probe compensator amplitude
g.
Check against limits:
H
Subtract the value just obtained (base level) from that obtained
previously (top level).
H
CHECK that the difference obtained is within 990 mV to 1010 mV,
inclusive.
H
Enter voltage difference on test record.
3.
Disconnect the hookup:
Disconnect the cable from
CH 1
.
Summary of Contents for TDS7054
Page 4: ......
Page 14: ...Table of Contents x TDS7104 TDS7054 Service Manual...
Page 18: ...Service Safety Summary xiv TDS7104 TDS7054 Service Manual...
Page 22: ...Preface xviii TDS7104 TDS7054 Service Manual...
Page 56: ...Operating Information 2 12 TDS7104 TDS7054 Service Manual...
Page 60: ...Theory of Operation 3 4 TDS7104 TDS7054 Service Manual...
Page 64: ...Performance Verification 4 4 TDS7104 TDS7054 Service Manual...
Page 76: ...Brief Procedures 4 16 TDS7104 TDS7054 Service Manual...
Page 86: ...Performance Tests Semi Automated Method 4 26 TDS7104 TDS7054 Service Manual...
Page 144: ...Performance Tests 4 84 TDS7104 TDS7054 Service Manual...
Page 160: ...Maintenance 6 6 TDS7104 TDS7054 Service Manual...
Page 230: ...Troubleshooting 6 76 TDS7104 TDS7054 Service Manual...
Page 232: ...Repackaging Instructions 6 78 TDS7104 TDS7054 Service Manual...
Page 238: ...Electrical Parts List 8 2 TDS7104 TDS7054 Service Manual...