Clear and Sanitize Procedures
10.
Push the Action option button until Save Waveform is selected. Push the Save To option button until Ref is selected.
Push the Source option button until CH1 is selected.
11.
Push the To option button until RefA is selected. Push the Save option button to overwrite reference waveform RefA
with the null waveform.
12.
Push the To and the Save option buttons again to overwrite the next reference waveform with the null waveform.
Repeat this step until reference RefA is selected again.
13.
For TDS oscilloscopes, insert a USB
fl
ash drive into the USB Flash Drive port on the front of the oscilloscope. For
TPS oscilloscopes, insert a compact
fl
ash card into the Compact Flash slot on the front of the oscilloscope.
14.
Push the front panel UTILITY button.
15.
Push the Options option button.
16.
Push the Printer Setup option button.
17.
Push the PRINT Button option until Saves Image To File is selected.
18.
For TDS oscilloscopes, push the File Format option button until JPG is selected. For TPS oscilloscopes, select
BMP instead of JPG.
19.
Remove any probes from the BNC front panel connectors.
20.
Push the front panel PRINT button to overwrite the hard copy image
fi
le with an image that does not contain any
useful information.
21.
Power off the oscilloscope, and then power on the oscilloscope to complete the process.
4
TDS1000B/TDS2000B, TDS1000C-EDU/TDS2000C, and TPS2000B Series