background image

Clear and Sanitize Procedures

Memory Devices

The following tables list the volatile and nonvolatile memory devices in the standard instrument and listed options:

Table 1: Volatile memory devices

Type and min.
size

Function

Type of user

information
stored

Backed up
by battery

Method of
modi

fi

cation

Data input
method

Location

User
accessible

To clear

To sanitize

SDRAM,
512K X 32

Acquisition memory
for holding
and processing
waveforms, and
processor system
RAM.

Both user data and
user settings

No

Indirect

Firmware
operations,
user input

Main
Acquisition
board

No

Remove power from
the instrument for at
least 20 seconds.

Remove power from
the instrument for at
least 20 seconds.

SRAM, 128K X 8

Shared memory
between the
acquisition system
and the USB
processor.

Both user data and
user settings

No

Indirect

Firmware
operations

Main
Acquisition
board. TDS
models only.

No

Remove power from
the instrument for at
least 20 seconds.

Remove power from
the instrument for at
least 20 seconds.

Real-time clock,
which contains
battery backed-up
memory

Holds date/time data

None

Yes

Indirect

User input

Main
Acquisition
board.

No

Remove the battery
from the board.

Remove the battery
from the board.

USB 2.0 OTG
host/peripheral
controller, 8K X 16
RAM

Contains USB
controller program
and data memory.

None

No

Indirect

Firmware
operations

Main
Acquisition
board. TDS
models only.

No

Remove power from
the instrument for at
least 20 seconds.

Remove power from
the instrument for at
least 20 seconds.

SRAM, 64K X16

Data memory for the
USB processor.

None

No

Indirect

Firmware
operations

Main
Acquisition
board. TDS
models only.

No

Remove power from
the instrument for at
least 20 seconds.

Remove power from
the instrument for at
least 20 seconds.

2

TDS1000B/TDS2000B, TDS1000C-EDU/TDS2000C, and TPS2000B Series

Summary of Contents for TDS1001B

Page 1: ...TDS1000B TDS2000B TDS1000C EDU TDS2000C and TPS2000B Series Digital Storage Oscilloscope ZZZ Declassification and Security Instructions xx www tektronix com 077 0445 00...

Page 2: ...and pending Information in this publication supersedes that in all previously published material Specifications and price change privileges reserved TEKTRONIX and TEK are registered trademarks of Tek...

Page 3: ...Preface iii Clear and Sanitize Procedures 1 Memory Devices 2 Data Export Devices 5 Troubleshooting 7 How to Clear or Sanitize a Nonfunctional Instrument 7 TDS1000B TDS2000B TDS1000C EDU TDS2000C and...

Page 4: ...Table of Contents ii TDS1000B TDS2000B TDS1000C EDU TDS2000C and TPS2000B Series...

Page 5: ...rument If you have any questions contact the Tektronix Technical Support Center at www tektronix com support Reference The procedures in this document are written to meet the requirements specified in...

Page 6: ...nual name Tektronix part number TDS1000B and TDS2000B Series Digital Storage Oscilloscope Service Manual 071 1828 XX TDS2000C and TDS1000C EDU Series Digital Storage Oscilloscope Service Manual 077 04...

Page 7: ...ll supplied to the instrument For the purpose of clearing data putting the instrument in Standby mode does not qualify as powering off For these products you will need to either push a rear panel OFF...

Page 8: ...Preface vi TDS1000B TDS2000B TDS1000C EDU TDS2000C and TPS2000B Series...

Page 9: ...stem resources modify the data The user cannot modify the data User accessible This column indicates whether the user can retrieve the device contents Yes The user can directly retrieve the memory dev...

Page 10: ...gs No Indirect Firmware operations Main Acquisition board TDS models only No Remove power from the instrument for at least 20 seconds Remove power from the instrument for at least 20 seconds Real time...

Page 11: ...in user data or settings Cleaning would disable instrument functionality Not applicable does not contain user data or settings Sanitizing would disable instrument functionality Clear Flash Procedure T...

Page 12: ...e front of the oscilloscope For TPS oscilloscopes insert a compact flash card into the Compact Flash slot on the front of the oscilloscope 14 Push the front panel UTILITY button 15 Push the Options op...

Page 13: ...strument setups Directly Save Front panel Files can be deleted or overwritten on the oscilloscope or a PC USB flash drive can be removed and destroyed The USB host port cannot be disabled USB device p...

Page 14: ...Clear and Sanitize Procedures 6 TDS1000B TDS2000B TDS1000C EDU TDS2000C and TPS2000B Series...

Page 15: ...regarding handling or disposal of the board Compact Flash Card TPS oscilloscopes only Remove the compact flash card and return the instrument to Tektronix for repair After removal of the compact flas...

Reviews:

Related manuals for TDS1001B