Clear and Sanitize Procedures

Table 2: Nonvolatile memory devices

Type and
minimum

size

Function

Type of user
information

stored

Method of

modi

fi

cation

Data input method

Location

User

Accessible

To clear

To sanitize

FLASH,
8M X 8,
4M X 16

Holds instrument

fi

rmware, current

setup, saved setups,
saved reference
waveforms, hard
copy image

fi

les, and

calibration constants.

Both user data and
user settings

Indirect

Factory con

fi

gurations,

fi

rmware operations,

user input, and USB disk
upgrade

Main
Acquisition
board

No

See the

Clear Flash

Procedure

that

follows this table.

See the

Clear Flash

Procedure

that follows

this table.

EEPROM,
SERIAL, 1M

Program memory for
the USB processor.

None

Indirect

Factory con

fi

gurations,

USB disk upgrade

Main
Acquisition
board. TDS
models only.

No

Not applicable,
does not contain
user data or
settings. Cleaning
would disable
instrument
functionality.

Not applicable, does
not contain user
data or settings.
Sanitizing would disable
instrument functionality.

Clear Flash Procedure

This procedure does not erase or change factory calibration constants.

1.

Push the front panel DEFAULT SETUP button to recall the default setup.

2.

Push the front panel SAVE/RECALL button.

3.

Push the Action option (side bezel) button until Save Setup is selected.

4.

Push the Save To option button until Setup is selected.

5.

Push the Setup option button until 1 is selected.

6.

Push the Save option button to overwrite setup 1 with the default setup.

7.

Push the Setup and Save option buttons again to overwrite the next setup with the default setup. Repeat this step until
setup 1 is selected again.

8.

Push the front panel CH 1 MENU button, and then push the Coupling option button until Ground is selected.

9.

Push the front panel SAVE/RECALL button.

TDS1000B/TDS2000B, TDS1000C-EDU/TDS2000C, and TPS2000B Series

3

Summary of Contents for TDS1001B

Page 1: ...TDS1000B TDS2000B TDS1000C EDU TDS2000C and TPS2000B Series Digital Storage Oscilloscope ZZZ Declassification and Security Instructions xx www tektronix com 077 0445 00...

Page 2: ...and pending Information in this publication supersedes that in all previously published material Specifications and price change privileges reserved TEKTRONIX and TEK are registered trademarks of Tek...

Page 3: ...Preface iii Clear and Sanitize Procedures 1 Memory Devices 2 Data Export Devices 5 Troubleshooting 7 How to Clear or Sanitize a Nonfunctional Instrument 7 TDS1000B TDS2000B TDS1000C EDU TDS2000C and...

Page 4: ...Table of Contents ii TDS1000B TDS2000B TDS1000C EDU TDS2000C and TPS2000B Series...

Page 5: ...rument If you have any questions contact the Tektronix Technical Support Center at www tektronix com support Reference The procedures in this document are written to meet the requirements specified in...

Page 6: ...nual name Tektronix part number TDS1000B and TDS2000B Series Digital Storage Oscilloscope Service Manual 071 1828 XX TDS2000C and TDS1000C EDU Series Digital Storage Oscilloscope Service Manual 077 04...

Page 7: ...ll supplied to the instrument For the purpose of clearing data putting the instrument in Standby mode does not qualify as powering off For these products you will need to either push a rear panel OFF...

Page 8: ...Preface vi TDS1000B TDS2000B TDS1000C EDU TDS2000C and TPS2000B Series...

Page 9: ...stem resources modify the data The user cannot modify the data User accessible This column indicates whether the user can retrieve the device contents Yes The user can directly retrieve the memory dev...

Page 10: ...gs No Indirect Firmware operations Main Acquisition board TDS models only No Remove power from the instrument for at least 20 seconds Remove power from the instrument for at least 20 seconds Real time...

Page 11: ...in user data or settings Cleaning would disable instrument functionality Not applicable does not contain user data or settings Sanitizing would disable instrument functionality Clear Flash Procedure T...

Page 12: ...e front of the oscilloscope For TPS oscilloscopes insert a compact flash card into the Compact Flash slot on the front of the oscilloscope 14 Push the front panel UTILITY button 15 Push the Options op...

Page 13: ...strument setups Directly Save Front panel Files can be deleted or overwritten on the oscilloscope or a PC USB flash drive can be removed and destroyed The USB host port cannot be disabled USB device p...

Page 14: ...Clear and Sanitize Procedures 6 TDS1000B TDS2000B TDS1000C EDU TDS2000C and TPS2000B Series...

Page 15: ...regarding handling or disposal of the board Compact Flash Card TPS oscilloscopes only Remove the compact flash card and return the instrument to Tektronix for repair After removal of the compact flas...

Reviews:

Related manuals for TDS1001B