Functional Overview

ST2400A SDH Test Set User Manual

2-11

Front Panel Connections

155 Electrical Add Input / 2.048 Ext Ref Clock Input (BNC)

This BNC connector is mounted on front panel 155e Input of all configurations. This connector is
only used in Transceiver and Transmitter configurations.

155 Electrical Drop Output (BNC)

This BNC connector is mounted on front panel 155e Output of all configurations. This connector
is only used in Transceiver and Receiver configurations.

155/622 Optical Add Input (FC, SC, or ST)

An optical connector bulkhead plus cap and chain is mounted on front panel 155/622 Optical Input
of all configurations. Bulkhead type (FC, SC, or ST) depends on customer-specified optical
connector type (required customer option). 155/622 Add optics will be installed and fiber
connected to this bulkhead only in TX/RX and TX Only configurations.

155/622 Optical Drop Output (FC, SC, or ST)

An optical connector bulkhead (customer-specified as FC, SC, or ST) plus cap and chain is
mounted on front panel 155/622 Optical Output of all configurations. 155/622 Drop optics will be
installed and fiber connected to this bulkhead only in TX/RX and RX Only configurations.

TX 2.4 Gb/s Optical Output (FC, SC, or ST)

An optical connector bulkhead (customer-specified as FC, SC, or ST) plus cap and chain is
mounted on front panel 2.4 Gb/s Output of all configurations.  No fiber is connected to this
bulkhead in RX Only configurations.  Optical connector from 2.4 Gb/s TX laser(s) is attached to
this bulkhead only in TX/RX and TX Only configurations when 1 or 2 optional lasers are ordered.

RX 2.4 Gb/s Optical Input (FC, SC, or ST)

An optical connector bulkhead (customer-specified as FC, SC, or ST) plus cap and chain is
mounted on front panel 2.4 Gb/s Input of all configurations.  No fiber is connected to this bulkhead
in TX Only configurations. Optical connector is fiber-connected to APD RX only in TX/RX and
RX Only configurations when optional APD RX is installed.

Summary of Contents for ST2400A

Page 1: ...User Manual p p S T 2 4 0 0 A 2 4 Gb s SDH Test Set 071 0241 04 This document supports firmware version 5 4 and above Advanced Test Equipment Rentals www atecorp com 800 404 ATEC 2832 Established 1981...

Page 2: ...ph c 1 ii of the Rights in Technical Data and Computer Software clause at DFARS 252 227 7013 or subparagraphs c 1 and 2 of the Commercial Computer Software Restricted Rights clause at FAR 52 227 19 as...

Page 3: ...oducts returned to any other locations This warranty shall not apply to any defect failure or damage caused by improper use or improper or inadequate maintenance and care Tektronix shall not be obliga...

Page 4: ...ications hardware and software by expert application engineers If you have any questions regarding the operation maintenance repair or application of your Tektronix equipment contact your local sales...

Page 5: ...2 4 Transmitter Connectors 2 4 Transmitter Status Indicators 2 5 Mode Key 2 5 Tributary and Input Select Keys Transmitter 2 5 OVHD Key Transmitter 2 6 Alarm Key Transmitter 2 6 Error Key Transmitter 2...

Page 6: ...neration Menus SDH 3 20 Overhead Menus SDH Transmitter 3 22 Ability to Edit Multi byte J0 Trace 3 23 J0 J1 Trace Capture Display 3 26 Ability to Edit M1 Byte 3 28 Error Generation Menus SDH 3 29 B2 or...

Page 7: ...ST2400A SDH Test Set User Manual vii Index I 1...

Page 8: ...y the power cord specified for this product Avoid Electric Overload To avoid electric shock or fire hazard do not apply a voltage to a terminal that is outside the range specified for that terminal Gr...

Page 9: ...ve 10dBm may damage the Optical Input devices Always pad the input level to less than 10dBm Always use 15 dB of attenuation when connecting the ST2400A Transmitter output to its receiver input Failure...

Page 10: ...ANGER indicates an injury hazard immediately accessible as you read the marking WARNING indicates an injury hazard not immediately accessible as you read the marking CAUTION indicates a hazard to prop...

Page 11: ...ntions This manual uses the following conventions The names of front panel controls and menus appear in all upper case letter for example TRANSMIT and HELP Names appear in the same case in this manual...

Page 12: ...General Safety Summary xii ST2400A SDH Test Set User Manual...

Page 13: ...Getting Started...

Page 14: ......

Page 15: ...APERTURE CLASS 3B LASER PRODUCT 21 CFR 1040 10 APR 97 OVHD MODE TRIB ALARM ERROR 10dBm MAX 0dBm NOM 10dBm MAX 8dBm MAX 155 622 INPUT STATUS LOS LOF INPUT SELECT 155 622 155 155 622 2 4 Gb s STATUS DRO...

Page 16: ...band optics 42 Add B3 Path Analysis ST2400T STM 16 OC 48 SDH SONET Transmitter Includes 2 488 Gbit s Transmitter with electrical clock and data tributary add power cord user manual 13 Add 1310 nm inte...

Page 17: ...eive and demultiplex the STM 16 signal and drop either the STM 1 or STM 4 signal to an auxiliary test set as required The ST2400A can accept zero one or two factory installed 2 488 Gb s laser transmit...

Page 18: ...e recovery mode Electrical Offset of Receiver Decision Threshold 1 nm wavelength tuning CW laser source mode 2 488 Gb s electrical I O 3 year warranty Automatic Optical Receiver shut off if power exce...

Page 19: ...Getting Started ST2400A SDH Test Set User Manual 1 5 CCITT ITU G 708 G 709 and GR 253 CORE framing...

Page 20: ...C2 byte can be set to any value from 0x00 to 0xFF B3 Path Analysis J1 Path Trace displayed as ASCII text Tx Patterns PRBS 2 31 1 and 2 31 1 Inverted Fixed 00 to FF TX mode button Pointers Read only m...

Page 21: ...e included with the ST2400A 2 4 Gb s SDH SONET Test Set If you wish to purchase optional accessory contact your local Tektronix Representative Standard User Manual Certificate of Traceable Calibration...

Page 22: ...ale to N 015 0572 00 SMA male to BNC male 015 0554 00 SMA male to BNC female 015 1009 00 SMA male to N female 015 0553 00 Threaded female to male slip on connector 015 0549 00 Male to female connector...

Page 23: ...the ST2400A which is exhausted through the top vents 4 Connect the unit for a chosen application Never exceed the maximum optical power limits 10dBm listed on the front panel When not using always co...

Page 24: ...Operating Basics...

Page 25: ......

Page 26: ...s in the ST2400A SDH are Class IIIA per IEC 825 1 TRIB OVHD ERROR 155 2 048 CLK IN 75 OHM ST2400A 2 4 Gb s SDH SONET Test Set GPIB LLO REM INC ENTER RUN STOP DEC SETUP PAUSE RESUME LOS B1 AIS LOF B2 R...

Page 27: ...est Set INC ENTER RUN STOP DEC SETUP PAUSE RESUME AVOID EXPOSURE TO BEAM INVISIBLE LASER RADIATION IS EMITTED FROM APERTURE CLASS 3B LASER PRODUCT 21 CFR 1040 10 APR 97 Figure 2 2 ST2400A LCD Display...

Page 28: ...TIME ENTER Figure 2 3 Examples of SDH Setup Main Menus Since the ST2400A can be configured as Transceiver Transmitter or Receiver many menus are available only with the appropriate configuration For...

Page 29: ...r internal signal generation using external timing synchronization SETS input accepts signals meeting the requirements of G 703 section 10 2 048 Mb s and section 6 2 048 Mhz with the exception of freq...

Page 30: ...the Transmitter front panel INT an internally generated 2 4 Gb s signal and THRU a through input from the Receiver section There are also choices for Pattern Selection Mapping Selection and External...

Page 31: ...detail is provided in Chapter 3 Reference RS EOW E1 PASS SET RS EOW E1 PASS SET 7F Main Menu Field Change Figure 2 5 Overhead Menu Access Alarm Key Transmitter The Transmitter alarm key is used to in...

Page 32: ...l connector is an industry standard BNC that is marked 155 OUT It provides the 75 ohm load impedance and direct coupling DC for the SDH electrical signal The two industry standard FC PC optical connec...

Page 33: ...that more than 3 milliseconds of an out of frame condition has occurred The OOF Out Of Frame alarm indicates that more than 4 consecutive frames of the 2 4 Gb s signal had frame alignment errors The...

Page 34: ...indicates a history condition an error of this type occurred at some point during the test but is not occurring at this time B3 h PL BIT P REI LOP P AIS h P RDI Figure 2 8 VIEW HISTORY additional LED...

Page 35: ...out effecting counter values The green LED indicator inside the key is on during the RUN cycle A stopped test can be restarted using the pause run resume sequence saving stored history data The button...

Page 36: ...p Output FC SC or ST An optical connector bulkhead customer specified as FC SC or ST plus cap and chain is mounted on front panel 155 622 Optical Output of all configurations 155 622 Drop optics will...

Page 37: ...ng remotely operated Rear Panel Description This section describes the rear panel controls and connectors PARALLEL PRINTER GPIB ST112 RS232 RS232 LINE 5AT 250V 2 5AT 250V FUSE SLOBO SLOBO MAX 250W 250...

Page 38: ...ts Receiver input Failure to do so will damage the optical detector SYNC Output SMA SYNC Output Clock Trigger is an AC coupled output The voltage level is 300 mV peak to peak minimum This output requi...

Page 39: ...ns 10 The pulse repetition rate is equal to the 125 The ST2400A provides single ended 2 488 Gb s clock and data interface inputs and outputs mounted on the rear panel Electrical clock and data output...

Page 40: ...oring applications It is perfect for LEC IXC local exchange carrier inter exchange carrier cross boundary or other long term passive performance monitoring It does not allow optical line error injecti...

Page 41: ...test In this application the CTS 850 or CTS 750 test set can provide the 155 Mb s or 622 Mb s framed signal input to the ST2400A which is actively inserted into the signal path The ST2400A provides a...

Page 42: ...the additional tributary test set to quickly test monitor and verify the 2 4 Gb s signal operation ST2400A 2 4 Gb s SDH SONET Test Set 2 4 Gb s ADM CTS 710 NETWORK OC 3 or OC 12 OC 48 OC 48 OC 48 Fig...

Page 43: ...ndication BER BER measurements based on B1 and B2 errors With concatenated signals the H1 and H2 byte 1 locations contain the location of the payload start container All other H1 and H2 byte locations...

Page 44: ...Reference...

Page 45: ......

Page 46: ...values The Enter key is also used to access submenus The Main Menus that make up the Setup menu are shown in Figure 3 1 The Main Menus are a continuous loop that begins with the Quick Setup LCD displa...

Page 47: ...PEAT REPORT SETUP SQUELCH OFF ON PRINTER SELECTION PARALLEL RS232 REPORT SETUP CLEAR ERR NO YES TEST TIME DD HH MM SS 00 00 00 00 REPORT SETUP PRINT OFF ERR EOT BOTH STORE RECALL SETUP USER FACTORY DE...

Page 48: ...LF CR CR LF AUXILIARY SETUP ENTER RESUME OPER ON RESTART OFF ON TIME YY MM DD HH MM SS 1998 07 22 14 40 32 GPIB ADDR 31 OFF BUS ADDR 14 TERM EOI LF SYSTEM CONFIGURATION SONET SDH ST2400A VERSION X XX...

Page 49: ...rate When the choice is made by pressing the Enter key alarms and errors if on will be turned off Overhead byte values will be reset and set to pass mode The next menu will be automatically displayed...

Page 50: ...253 LR 1 G 957 L 16 1 1550nm DM GR 253 LR 2 G 957 L 16 2 To advance to the next LASER MODULE if one exists press the SETUP key on the front panel The SETUP and CW choices will not be available for ILM...

Page 51: ...less than 1 s It can be used to synchronize other equipment such as an oscilloscope trigger source The TX Event Trigger Output menu is available in the Transceiver or Transmitter configurations TX EVE...

Page 52: ...mory locations 0 through 9 A unique 10 character label can be stored with each saved setup The arrow keys are used to position the cursor anywhere in the label area and the INC or DEC key is used to s...

Page 53: ...Reference 3 8 ST2400A SDH Test Set User Manual time set in the TEST TIME menu then start the test again The stop key can stop any test that is running...

Page 54: ...on The error log should be cleared before running a long test preventing buffer overflow An error log warning message will be displayed when the run key is pressed if the log is 95 full 7 The REPORT R...

Page 55: ...ser to turn on and the device under test DUT to settle Before this change after a loss of power a measurement would need to be restored due to the 2 4 Gb s laser being turned off If System Configurati...

Page 56: ...B2 B3 Payload Bit FEBE P Rath RDI and FEBE L MS REI errors In SDH mode the ST2400A will provide Errored Blocks EB and Background Block Errors BBE for G 826 error analysis measurement In SDH mode the...

Page 57: ...r panel Sync output signal characteristics AC coupled 300 mV pk pk Terminate externally with 50 Ohms Pointers Read only menu under RX Trib button Pointer Value displayed 0 to 782 Pointer Measurement C...

Page 58: ...ations will not include 155 622 optical and electrical drop interface although unused connectors may be mounted Receiver configurations will not include 155 622 optical and electrical add interface al...

Page 59: ...r Log becomes full If the Error Log is more than 95 full when the RUN key is pressed the message below will be displayed Selecting NO will allow the test to run but when the error log is full new data...

Page 60: ...tus section there are three yellow LED indicators two optical and one electrical The appropriate LED will indicate the input signal The INT mode generates an unequipped SDH 2 4 Mb s STM 16 signal with...

Page 61: ...2 4 Gb s SDH signal that entered the Receiver from an external source All three yellow status LEDs will be off when the ST2400A is in the THRU mode In older ST2400 test sets pressing the MODE key a se...

Page 62: ...pear when the INTERNAL Transmitter mode is enabled A Loss of Signal message will be displayed on the front panel LCD if either of these options are selected and the signal is not present If the 2 048...

Page 63: ...nally generated 2 4 Gbit s signal to which externally supplied 155 or 622 Mbits signal is added EXT mode The user has the same background mapping selections STS1 or STS3c in SONET VC4 in SDH Payload a...

Page 64: ...Mb s TRIB INSERT SINGLE n ALL Figure 3 11 Transmitter Tributary Select Menus A choice is offered between multiplexing the tributary into one specific channel or into all available 2 4 Gb s channels Th...

Page 65: ...XT MUST BE SELECTED Figure 3 12 Transmitter Alarms If the TRIB or INPUT SELECT keys are pressed when the Transmitter is in the INT internal or THRU through mode the illegal mode display will be exhibi...

Page 66: ...ntered at the bottom LOS Loss Of Signal ERRORED FRAME MS RDI MS AIS and INJECT The alarm menus are used to set specific parameters simulating various synchronization defects as detailed below Each of...

Page 67: ...g the alarm key to leave the menu will not turn the alarm off The INJECT xxxx alarm menu permits recurring alarm conditions to be set Position the cursor under the alarm xxxx in this case and use the...

Page 68: ...used to pass or change the 2 4 Gb s signal overhead bytes in the first column of the STM 1 The Transmitter OVHD overhead key is used to scroll through the Regenerator Section RS and Multiplexer Secti...

Page 69: ...ecent List Report GR 253 ILR Issue 2B Dec 97 Some vendors are promoting 64 byte trace matching SONET J1 trace standard in GR 253 CORE ST2400A ability to generate non standard compliant 16 byte and 64...

Page 70: ...rmatted trace and 64 byte trace as shown below J0 TRACE SELECTION 16 BYTE 64 BYTE When 16 byte trace is selected 15 editable bytes of 16 byte trace are set to first 15 bytes of trace in memory If 16 b...

Page 71: ...s hex 00 and CR LF is placed in last two characters When trace characters are selected left arrow right arrow INC DEC and ENTER are used to navigate trace and edit characters If trace contains non pri...

Page 72: ...received No CRC error detection capability is provided in 16 byte trace mode 64 byte trace selection expects 64 byte trace matching format specified for STS Path Trace in GR 253 CORE as described abo...

Page 73: ...perly updated 16 BYTE EDIT CLR HEX 54 TEK ST2400A CRC F4 When 64 Byte is selected as shown below first 22 characters following CR LF is displayed as shown Non printable characters is displayed as boxe...

Page 74: ...d to scroll through the Regenerator Section RS and Multiplexer Section MS overhead bytes The top line of each display identifies the designated SDH overhead byte The bottom line is used to either PASS...

Page 75: ...r will disable all other active errors The first menu SINGLE ERROR INJ is used to inject one error each time the Enter key is pressed Use the arrow keys to positioning the cursor over the B1 B2 B3 MS...

Page 76: ...ey to move the cursor to OFF and press ENTER The asterisk and the word ACTIVE will no longer be displayed indicating the error injection is off The RANDOM BER Bit Error Rate INJECT main menu is used t...

Page 77: ...payload or the user an inject B1 or B2 errors simultaneously in all channels of 2 4 Gbit s payload B2 error rates may be generated over the range 1 0E 3 to 0 1E 9 SINGLE ERROR INJECT B1 B2 B3 MS REI T...

Page 78: ...l and press the ENTER key to generate that error When the cursor is positioned over B1 press the INC key again to change to the B2 byte The next menu permits injecting a single B2 error into all chann...

Page 79: ...and again to cycle through the choices RS B1 Measurement RS ES 0 BER 1 0e 04 B1 SES 0 TOT 0 The Enter key will flash while at the menu above Pressing the Enter key permits access to the following sub...

Page 80: ...ES 0 BER 1 0e 04 B2 SES 0 TOT 0 The Enter key will flash while at the menu above Pressing the Enter key permits access to the following sub menus MS EB 0 CSES 0 B2 BBE 0 MS BBER ESR 0 0E 00 B2 0 00E 0...

Page 81: ...ES 0 BER 1 0e 04 B3 SES 0 TOT 0 The Enter key will flash while at the menu above Pressing the Enter key permits access to the following sub menus HP EB 0 CSES 0 B3 BBE 0 HP BBER ESR 0 0E 00 B3 0 00E 0...

Page 82: ...0 BER 1 0e 04 REI SES 0 TOT 0 The Enter key will flash while at the menu above Pressing the Enter key permits access to the following sub menus MS EB 0 CSES 0 REI BBE 0 MS BBER ESR 0 0E 00 REI 0 00E...

Page 83: ...0 BER 1 0e 04 REI SES 0 TOT 0 The Enter key will flash while at the menu above Pressing the Enter key permits access to the following sub menus HP EB 0 CSES 0 REI BBE 0 HP BBER ESR 0 0E 00 REI 0 00E...

Page 84: ...TSE ES 0 BER 1 0e 04 SES 0 TOT 0 The Enter key will flash while at the menu above Pressing the Enter key permits access to the following sub menus TSE EB 0 CSES 0 BBE 0 TSE BBER ESR 0 0E 00 0 00E 00...

Page 85: ...a set of submenus for each of the first three main menus RS B1 MS B2 HP B3 MS REI HP REI and TSE Measurement Errors that were detected on the 2 4 Gb s SDH signal are displayed in the RS B1 MS B2 and H...

Page 86: ...SR Severely Errored Seconds Ratio conforms to the ITU G 826 standard The submenu AEB ABBE AES and ASES are the Errored Blocks Background Block Errors Errored Seconds and Severely Errored Seconds count...

Page 87: ...Errored Second Ratio The ratio of Errored Seconds ES to total seconds in available time since testing was started or restarted SESR Severely Errored Second Ratio The ratio of Severely Errored Seconds...

Page 88: ...00 00 00 00 RX OH B1 E2 F1 B3 00 00 00 00 RX OH D10 D11 D12 Z4 00 00 00 00 LOS DURATION 2422ms LOF DURATION 2428ms JO TRACE SELECTION 16 BYTE 64 BYTE J1 TRACE SELECTION 16 BYTE 64 BYTE Figure 3 19 Rec...

Page 89: ...trace capability ASCII Synchronization Messages S1 byte For convenience both the SONET and SDH modes are compared below The S1 byte is in hexadecimal The unused SONET values are left blank The unknow...

Page 90: ...the line requesting to be switched and the Switch Line the line that the Request Line wants to switch to can be any decimal number from 0 to 15 APS 1 1 RQ 0 SW 0 0 No Request Direction Protection Requ...

Page 91: ...front panel increment and decrement keys This display depends on which drop was selected in the RX DROP SELECT or SDH QUICK SETUP menu off 155 or 622 The two possible SDH menus are shown below 155 Mb...

Page 92: ...eiver mappings and channels to the ones available from the ST2400A receiver If an invalid mapping is detected a menu will flash on the LCD screen of the test unit When a valid mapping is detected the...

Page 93: ...or and disable the optical drop The 155 Mb s STM 1 optical drop or the 622 Mb s STM 4 optical drop can be selected by choosing either the 155 or 622 respectively The 155 or 622 LED on the Receiver sec...

Page 94: ...Reference ST2400A SDH Test Set User Manual 3 49...

Page 95: ...Appendices...

Page 96: ......

Page 97: ...ximum Fuses The AC power module is located in the lower left hand corner of the rear panel It has a removable fuse panel that can be opened with a small flat blade screw driver after the AC Line Cord...

Page 98: ...Standard EC Declaration of Conformity Compliance was demonstrated to the following specification as listed in the Official Journal of the European Communities EC Council Directive 89 336 EC AS NZS 20...

Page 99: ...ut so no pointer adjustments are required or generated 155 All Replicate Mode Data 2 488 Gb s output is made from 16 copies of the user selected incoming 155 Mb s optical or electrical data 155 Single...

Page 100: ...is made with 16 channels of unequipped VC4 structured STM 1s bulk filled with the user selected test pattern all zeros PN23 inverted PN23 ITU T O 181 test pattern PN31 or inverted PN31 Default Overhe...

Page 101: ...p J0 E1 F1 D1 D3 group selected C2 K1 K2 D4 D12 group selected S1 M1 E2 The following TOH bytes of the first STM 1 of the STM 16 frame CANNOT be edited A1 A2 B1 B2 H1 H3 J1 Some TOH bytes can be alter...

Page 102: ...ros for a user selected time from 1 to 127 microseconds with a resolution of 1 microseconds or 0 1 to 9 9 seconds or continuously LOF OOF All of the A1 and A2 bytes of the 2 488 Gb s output are invert...

Page 103: ...ms Detection LOS 2 488 Gb s input is monitored for the all zero condition When the 2 4 Gb s signal has been off for more than 2 microseconds the LOS condition is declared The LOS alarm will assert no...

Page 104: ...ed by the ST2400A H1 H2 and H3 are copied from the selected drop channels The remainder of the drop signal overhead is copied from the first three columns of the 2 488 Gb s input Output This mode is v...

Page 105: ...t 15 17 1544 1556 nm not including the laser temperature tuning Wavelength select 1550 nm Opt 1530 1565 nm per ITU G 692 grid Adjustment Range applies only to 1550 nm lasers 1 0 nm Adjustment Resoluti...

Page 106: ...trical Input Connector BNC 75 Type CMI per Bellcore GR 253 CORE and CCITT Rec G 708 G 709 CMI is adjusted and guaranteed to meet output pulse mask only at room temperature 25 degrees C 5 degrees C Des...

Page 107: ...rminated ECL requires connection to an AC coupled 50 Ohm termination through a bias tee network supplying current for the ECL outputs falling edge of output clock coincides with data edge crossing 100...

Page 108: ...Optical Input Power under bias no damage 10 dBm APD enabled Damage will occur if exceeded Pi Max Optical Input Power under no bias 10 dBm APD Disabled Damage will occur if exceeded Fmax Maximum data r...

Page 109: ...and data interfaces allow ST2400A to operate error free when electrical outputs are looped to electrical inputs through a bias tee termination data valid when sampled within 50 ps of center of data ey...

Page 110: ...scription 155 Mb s electrical data output from demultiplexer intended for connection to supplemental test equipment GPIB I O Connector signal level and protocol per IEEE 488 2 Description Byte serial...

Page 111: ...uency Range 2 48832 Gb s 20 ppm Internal Timing The output timing is obtained from an internal clock source This timing mode can be used if the generator is in internal mode operating stand alone Outp...

Page 112: ...88 GHz The Receiver divides its input 2 488 GHz clock to 155 MHz or 622 MHz for the tributary drop rates or to 78 MHz for the Transmitter in THRU Mode The resulting maximum peak peak Jitter Gain is sh...

Page 113: ...Specifications SDH ST2400A SDH Test Set User Manual A 17...

Page 114: ...last page of instructions for required equipment NOTE This procedure may also be used to verify correct operation of an ST2400A Transmit Only but a known good ST2400A Receiver is required Similarly a...

Page 115: ...e the default settings set the laser to OFF so the unit is no longer linked Now that the unit has been correctly defaulted you may continue with the next section of the procedure STM 16 SDH Quick Setu...

Page 116: ...the receiver There are 27 steps in this section of the procedure 1 On the transmitter side of the front panel press MODE until the display shows the INT TX VC4 TEST PATT menu 2 Using the LEFT RIGHT c...

Page 117: ...the display shows the Single Error Inject menu 16 Using the LEFT RIGHT cursors select MS REI and then press ENTER twice 17 In the 2 4Gb s STATUS section observe that the error light for REI FEBE flas...

Page 118: ...E mode and will freeze all error conditions When the red LED is within the PAUSE RESUME switch is turned off the unit will RESUME normal functioning 7 On the display press PAUSE RESUME until the red L...

Page 119: ...hat J0 equals 01 10 On the transmitter side of the front panel press OVHD until the display shows D7 D8 and D9 on the top right side 11 Using the LEFT RIGHT cursors select SET and select D7 12 Using t...

Page 120: ...the INC DEC keys and the LEFT RIGHT cursors change D9 to 00 and then press ENTER 26 On the receiver side of the front panel press OVHD until the display shows D7 D8 and D9 on the top right side 27 On...

Page 121: ...TATUS section observe that the LOS LOF OOF B1 B2 and REI FEBE LEDs are now lit 6 Using the LEFT RIGHT cursors select OFF and then press ENTER 7 In the 2 4Gb s STATUS section observe that all of the er...

Page 122: ...eates an AIS alarm and you verify the responses of the receiver There are eleven steps in this section of the procedure 1 On the transmitter side of the front panel press ALARM until the display shows...

Page 123: ...ront panel press ALARM until the display shows the Inject LOS menu 2 Using the LEFT RIGHT cursors select ENABLE and then press ENTER 3 In the 2 4Gb s STATUS section observe that the LOS LOF OOF B1 B2...

Page 124: ...nd verify that the unit responds correctly This section is split into two parts Single Error Inject Test and Error Rate Inject Test Leave the setup as it was in the last section Single Error Inject Te...

Page 125: ...Incoming Inspection Test B 12 ST2400A SDH Test Set User Manual 12 On the front panel press RUN STOP twice Observe that the errors have now been reset to zero...

Page 126: ...E until the red LED within the switch is no longer lit 8 On the display observe that the error rates are once again increasing Also in the 2 4Gb s STATUS section observe that the error lights for B1 B...

Page 127: ...9 On the display observe that J0 equals 01 10 On the transmitter side of the front panel press OVHD until the display shows D7 D8 and D9 on the top right side 11 Using the LEFT RIGHT cursors select S...

Page 128: ...or lights for AIS and RDI FERF may flash 6 Using the LEFT RIGHT cursors select OFF and then press ENTER 7 In the 2 4Gb s STATUS section observe that all of the error lights are now off 8 On the front...

Page 129: ...his part the transmitter creates an AIS alarm and you verify the responses of the receiver There are eleven steps in this section of the procedure 1 On the transmitter side of the front panel press AL...

Page 130: ...OF OOF B1 B2 and REI FEBE LEDs flash on and off continuously every 2 5 seconds 4 Using the LEFT RIGHT cursors select OFF and then press ENTER 5 In the 2 4Gb s STATUS section observe that all of the er...

Page 131: ...sh_type 0 MIN rx_err_thresh_value 3 rx_drop_rate RX_DROP_RATE_OFF rx_drop_chan 1 rx_laser_on OFF rx_sts_chan 0 rx_eot_print OFF rx_eot_print_form 0 rx_errlog_print 0 rx_errlog_squelch NONE rx_measure_...

Page 132: ...le_error_type OFF tx_channel 1 tx_chan_replicate_mode 0 tx_input_mode INTERNAL tx_wavelength 1310 or 1550 depending on DUAL Wavelength or NOT tx_input_rate OFF tx_optic_on OFF tx_event_trigger EVENT_T...

Page 133: ...eters menu_key 0 quick_set INT 155 setup_store_mode OFF setup_store_select 0 System Parameters NOT reset by rst gpib_term 0 rs232_baud 9600 rs232_baud_code 0 rs232_parity EVEN rs232_char_size 0 rs232_...

Page 134: ...Factory Defaults C 4 ST2400A SDH Test Set User Manual...

Page 135: ...and the relationships between each are shown in the two tables that follow Table F 1 SONET SDH Hierarchies SONET Bit Rate SDH SONET Capacity SDH Capacity STS 1 OC 1 51 84 Mbit s STM 0 28 DS1 or 1 DS3...

Page 136: ...Mbit s STM 16 2 4 Gbit s and STM 64 10 Gbit s have also been defined Multiplexing is accomplished by combining or interleaving multiple lower order signals 1 5 Mbit s 2 Mbit s etc into higher speed c...

Page 137: ...he remaining traffic passes straight through the multiplexer without additional processing Map Demap A term for multiplexing implying more visibility inside the resultant multiplexed bit stream than a...

Page 138: ...hich develops and publishes standards for voluntary use in the United States Asynchronous A network where transmission system payloads are not synchronized and each network terminal runs on its own cl...

Page 139: ...ne binary digit a pulse of data Bit Error vs Block Error Error rate statistics play a key role in measuring the performance of a network As errors increase user payload especially data must be re tran...

Page 140: ...ous data structures for example two channels joined to form a single channel In SDH a number M of TUs can be linked together to produce a concatenated container M times the size of the TU An example o...

Page 141: ...itute Organization responsible for defining and maintaining European standards including SDH Failure A termination of the ability of an item to perform a required function A failure is caused by the p...

Page 142: ...nput signals in an efficient manner thus allowing higher transmission rates ITU International Telecommunications Union An agency of the United Nations responsible for the regulation standardization co...

Page 143: ...ies with the transmit rate but corresponds to a 2 10 7 BER Locked Mode A virtual tributary mode that fixes the starting location of the VC Locked mode has less pointer processing than floating mode LO...

Page 144: ...ons such as locating the payload in the frame multiplexing or concatenating signals performance monitoring automatic protection switching and line maintenance Multiplexer A device for combining severa...

Page 145: ...channel in the same bitstream Parity check An error checking scheme which examines the number of transmitted bits in a block which hold the value of one For even parity an overhead parity bit is set...

Page 146: ...tification that a failure condition has been declared downstream RAI signals were previously referred to as Yellow signals Remote Defect Indication RDI A signal returned to the transmitting Terminatin...

Page 147: ...tical Network A standard for optical transport in the United States Canada Japan Korea and Hong Kong that defines optical carrier levels and their electrically equivalent synchronous transport signals...

Page 148: ...igher order path layer It contains the Virtual Container VC plus a tributary unit pointer Tributary Unit Group TUG Contains several Tributary Units UI Unit Interval a measure of jitter UIpp Unit Inter...

Page 149: ...tes ITU T G 703 Physical electrical characteristics of hierarchical digital interfaces G 704 Synchronous frame structures used at 1544 6312 2048 8448 and 4476 kbit s hierarchical levels G 706 Frame al...

Page 150: ...nts and systems relating to the synchronous digital hierarchy ITU T G 958 Digital line systems based on the Synchronous Digital Hierarchy for use on optical fibre cables I 432 B ISDN User network inte...

Page 151: ...3 32 F First Time Operation 1 8 Front Panel 2 1 Connections 2 10 Display Display Controls Indicators 2 2 Setup Menus 2 2 G General Menus 3 10 Error Log Display 3 13 Menu Measurement Refinements 3 10...

Page 152: ...lock 2 12 Remote Indicators GPIB 2 11 S Safety Summary ix SDH Glossary G 1 Setup Menus 3 1 Auxiliary Setup 3 8 Laser Menus 3 4 Power Fail Recovery 3 9 Report Test Time 3 7 Store Recall 3 7 TX Event Tr...

Reviews: