![Tektronix SC 502 Instruction Manual Download Page 47](http://html.mh-extra.com/html/tektronix/sc-502/sc-502_instruction-manual_1079081047.webp)
Calibration Procedure-SC 502
Performance
Check
5. Check the Vertical
Bandwidth (1
mVldlv)
VO L
TS/DIV (CH
1
and
CH 2)
1
m
a.
Return
the
generator
to
50
kHz and
adj ust
t he
l eveled
si ne-wave
g
e
n
e
r
a
t
o
r
for
a
crt di spl ay of
e
x
a
c
t
l
y
six
divisi ons.
b.
Set
the
generator frequency to
5
MHz and adj ust the
freq uency
v
a
r
i
a
b
l
e
unt i l the
di spl ayed signal
is
at
l east
4.2
major
div i sions
in
am pl itude.
c.
Check-that
the
f
r
e
q ue
n
c
y
readout from
the
generator
i s equal
to or
g
r
e
a
t
e
r
than
5.0
MHz.
d.
Change
the
M O D E
switch to
CH 2.
e.
Connect
the generator out put to
CH 2
i n put
connec
tor
and return
the
generator
fr
e
qu
e
n
c
y
to 50 kHz and
set
the
amplitude
for
six
divisi ons.
f.
Repeat
parts b and
c to
c
h
e
c
k
the
CH
2
bandwi dth.
g. Disconnect aU
c
a
b
l
e
s
and
return front
panel
con
trols
to
t
h
e
prelimi nary setti ngs.
6. Check the AC LF
Response
MODE
CH l
VO L
TS/DIV
(CH
1
and
CH
2)
DC-GN D-AC
SEC O N DS/DIV
.5
DC
AMPL
a.
Adj ust
the
INTEN
a
n
d
FOCUS
controls
for
a
wel l
defi ned
l i ne
on
the crt
display.
b.
Connect
a
10 Hz
s
i
n
e-
w
a
v
e signal from the
function
generator
th
r
o
u
g
h
a
50
n
coaxial
cable,
w
i
t
h
a
50 n
termi nati on,
to
t
h
e
CH
1
i nput
connector.
c.
Adj ust
the
fu nction
generator
output
am plitude
for
si x graticule
divisions
of
display
(one
vert
i
c
a
l
line
dis
pl ayed
)
.
d. Set the
DC-GN D-AC
switch
to
AC.
4-4
e.
Check-that
the
displayed
am plit ude
is
greater
t han
4.2
divisions.
f.
Set
the
M O D E
switch to
CH
2,
reconnect
the
generator
to
the
CH
2
i nput
connector
and
repeat parts b
through
e.
g.
Disconnect
all
c
a
b
l
e
s
and
ret urn
the
front- panel
controls
to
the
prelimi nary
setti ngs.
7.
Check
the Displayed Noise
VOL TS/DIV (CH
1
and
CH 2)
MODE
DC-GND-AC
SOURCE
SECONDS/DIV
1
m
CH
2
DC
EXT
5.0
iJS
a
.
Connect
a
0.2
mV
standard
am plitude
si gnal
from
the
cali bration
generator
through a
50 n
coaxial cabl e to
the
CH
2
i nput connector
.
b.
Adj ust
the
I NTEN
and
FOCU S
controls
for a
wel l
defi
n
e
d
di splay.
c.
C
h
e
c
k
-
to
disti nguish
space
between
the two
l i nes
on
the
crt disp
l
a
y
.
d.
Change
the
D I S PLAY
switch
to
CH
1
and
reconnect
the
generator
to
t
h
e
CH
1
i nput
connector
and
repeat
parts
a
through
c.
e.
Disconnect
a l l
c
a
b
l
e
s
and
r
et
u
r
n
the front-panel
c
o
n
t
r
ol
s
to
the
prel i m i nary
setti n gs.
8. Check the Position
Range
SECON DS/DIV
. 5
m
VOL TS/DI V
(CH
1
a
n
d
CH
2)
DC-GN D-AC
10
m
A
C
a.
Connect
a
0. 1 2
V
s
t
a
n
d
a
r
d
am plitude
signal
from
the
cal i bration
generator
through
a 50 n c
o
a
xi
a
l
cable
to
the CH
1
i nput connector.
b.
Rotate
the
vertical
POSITI ON
control
f ul l y
c
l
o
ck
w
i
s
e
and
counterclockwise
and note
the
positi on of
the
peaks
of
the
displayed
waveform .
REV 0 APR 1 980
Summary of Contents for SC 502
Page 1: ...SC 502 OSCILLOSCOPE ...
Page 2: ......
Page 9: ...SC 502 1878 1 The SC 502 Oscilloscope xi REV APR 1982 ...
Page 10: ...Verso Filler Page z z ...
Page 32: ...Verso Filler Page z z ...
Page 34: ...Verso Filler Page z z ...
Page 63: ...Section 6 SC 502 OPTIONS There are no options available at this time REV A APR 1980 6 1 ...
Page 84: ...Beam R873 C473 Ext Bal R535 CH 1 Vert Po R335 CH 2 Vert Pos R394 ADJUSTft TP980 5 ...
Page 125: ...REV MAR 1 984 ...
Page 126: ...SC 502 m m o ...