Operating
Inltructlonl-SC 502
5.
Adj ust
the
horizontal
Positi o n control
to
m
o
v
e
the
10010
poi nt
of
the waveform to
the second verti cai l i ne
on
the graticule.
6.
Measure the horizontal
distance
between
the
1 00/0
and
90%
pOints.
Be
sure
t h e vari able SECONDS/DI V
contrel
is
i n
the
ful ly
clockwi se
positi on.
7.
M ulti ply
the
distance
measured
in
step
6 by
t he
setti n g
of
the
SECONDS/DIV switch.
EXAM PLE:
Assume that the horizontal di stance
between
the
100/0
and 90%
p
o
i
n
t
s
is
four
d
iv
i
s
i
o
n
s
and
the
SECONDSIDI V
s
wi
t
c
h
is
set to
1
JIS.
Usi ng the
period
form ul a
to fi nd
risetime:
Rise
time
peri od
horizontal
=
distance
X
(divisions)
The
risetime
is
4 JIS
.
SECONDSI
DIV
switch
setti ng
Time Difference Measurements
=
(4) ( 1 p&) = 4J1S
The
SC
502
Oscilloscope
permits
measurement
of
time
differen
ce
between
two
or
more
separate
events.
To
measure time
difference,
use
the
foll owing procedure:
1.
Set
the AC-G ND-DC switches
of both
channels
t
o
either
AC
or
DC.
2.
Set
the MODE
switch
to
either
CHOP
or
ALT. I n
general,
Chop
i s
more
suitable
for low-frequency
si gnals.
More
information
on
determi ning
the
mode
i
s given
under
Verti cal
D
i
s
pl
a
y
s
in
this secti on.
3. Set
t
he triggeri
ng
switches
to
tri
gger
the
d
i
s
p
la
y
on
channel
1
or channel
2,
whi chever
channel
p
ro
v
i
d
e
s
the
most
stable and
correct
triggeri ng
for
both
channels.
4.
Connect the
reference
signal
to the
channel l
i n
p
u
t
connector
and
the comparison
s
i
g
n
a
l
to
the
channel
2
i nput
connector.
The
start of the
reference
Signal
shoul d
lead that
of the
comparison
signal.
Use
coaxial cables
or
probes
that have similar time-del ay
characteristi cs
to
c
o
n
n
e
c
t the
Signal
to
the
input
connectors.
5.
If - the
signals are
o
PP
o
s
i
t
e
pol arity,
take
this i nto
account
in
the
final
calculation.
2-12
6. Set
the
channel VOL
TS/DIV
switches
t
o
produce
about four
divisions
of
display waveform.
7. Set
the
triggeri ng
control s
for
a stable
display. Set
the
SECON DSIDI V
switch
for
a
sweep rate
whi ch
shows
three
or more
divisions
between the
measurement pOi nts,
if possi ble.
8.
Adj ust
the
channel POSI TION control s
to
bri ng
the
measurement
points
to the
center
horizontal
reference
li ne.
9.
Adj ust
the
horizontal
P
O
S
I
TI
O
N
control
so the
channel
1
waveform
(reference)
crosses
the
center
horizontal l i ne
at
a
vertical
graticule
li ne.
10. Measure the
horizontal
distance between the
two
measurement
pOints
(see
Fi g. 2-7).
1 1
.
M ulti ply
the
m
e
a
s
u
r
e
d di stance
by
the
s
e
tti
n
g
of
the
SECON DS/DIV
switch.
�
Iitude
vel
�
Amp
Ie
Ch.nnel 2
1878·3
Rg. 2-7. Me
••
urlng time dlHerenc. between two pUll
EXAM PLE:
Assume
that
the SECONDS/DIV
switch
is
set to
50
JIS and
the
horizontal
distance between
measu re
ment
points
is
four
divisions.
U
si
n
g
the formula:
Time
Delay
=
SECO N D SI
D I V
switch
setting
hori zontal
X
di stance
=
(50 JIS) (4) =200
JlS
(di visions)
The time
delay
Is
200
JIS.
@
MAY 1 980
Summary of Contents for SC 502
Page 1: ...SC 502 OSCILLOSCOPE ...
Page 2: ......
Page 9: ...SC 502 1878 1 The SC 502 Oscilloscope xi REV APR 1982 ...
Page 10: ...Verso Filler Page z z ...
Page 32: ...Verso Filler Page z z ...
Page 34: ...Verso Filler Page z z ...
Page 63: ...Section 6 SC 502 OPTIONS There are no options available at this time REV A APR 1980 6 1 ...
Page 84: ...Beam R873 C473 Ext Bal R535 CH 1 Vert Po R335 CH 2 Vert Pos R394 ADJUSTft TP980 5 ...
Page 125: ...REV MAR 1 984 ...
Page 126: ...SC 502 m m o ...