SPECIFICATION
Perfonnance Conditions
The etectriCal
characteristics in
this
specification
are
valid
with the
following conditions:
Any conditions that are uniQue to a particular Character.
istic
are
expressly stated as par1 of that characteristic.
1.
The
instrument must have
been
adjusted at
an
ambient temperature
between
+20·C
and
+30·C.
2.
The
instrument must
be
in a non-<:ondensing environ.
ment wnose
limits are
described under
Environmental.
3.
Allow
twenty
minutes
warm-up time for
oper
ation to
specified
accuracy;
sixty minutes after exposure to or
storage
in
a high humidity (CXlndensirlg) environment.
The
electrical
and
environmental petformsnce limits to
gether
with their
related validation procedures are a
com
plete statement
of the efectrical
and environmental
performance
of
a calibrated instrument.
Table
1.1
VERTICAL DEFLECTION
SYSTEM
Characteri.tica
Peftonnance
Requirement.
lupplementallnfomlatJon
0eftec:00n
Factor
Calibrated Range
1
mV/div
to 20
V/drv in
14
steps in a
1.2-5
sequence.
Variabie
Range
At
least
2.5
to
1.
ContinuOusly variabie between calibrated
steps and extends maximum uncalibrated
deftection
factor to at
least
50
V/div.
DC
Balance
0.5
div
+ 1
mV.
(+15·C
to
+3S·C)
Accuracy
+1S·C
to
+35·C
5
mV/div to
20
V/div
:t2%
1
mV Idiv to
2
mV/div
:t5%
O·C
to
+50-C
Derated
IJrUrICf by
an
additiOnal
1 %.
Unearity
0.1
div or
less
of
mrTIprBSsiorl or expanSiOn
as
a
2
div
signal
is
positioned
becY.-een
the
gratlaJle limits.
HF
BandWidth
5
mV/div
to
20
V Idiv
At least
15
MHz.
Rasetine (caIoJated)
23
ns or
less
.
Typical step
response
:t
2%. 3%
peak-to-peak or less. Signal
aberrations
limited to
S
major
dMsions.
2mV/div
At
least
10
MHz.
1
mV/div
At least
S MHz.
REV FEB 1982
,-,
Summary of Contents for SC 502
Page 1: ...SC 502 OSCILLOSCOPE ...
Page 2: ......
Page 9: ...SC 502 1878 1 The SC 502 Oscilloscope xi REV APR 1982 ...
Page 10: ...Verso Filler Page z z ...
Page 32: ...Verso Filler Page z z ...
Page 34: ...Verso Filler Page z z ...
Page 63: ...Section 6 SC 502 OPTIONS There are no options available at this time REV A APR 1980 6 1 ...
Page 84: ...Beam R873 C473 Ext Bal R535 CH 1 Vert Po R335 CH 2 Vert Pos R394 ADJUSTft TP980 5 ...
Page 125: ...REV MAR 1 984 ...
Page 126: ...SC 502 m m o ...