![Tektronix RSA6120A series Quick Start User Manual Download Page 70](http://html1.mh-extra.com/html/tektronix/rsa6120a-series/rsa6120a-series_quick-start-user-manual_1078170070.webp)
Operation
Using Correction Tables
The RSA6100A Real-Time Spectrum Analyzer lets you correct the signal displays and measurement results to account
for gains and losses in external equipment. You can specify a single gain value or you can use an external loss table to
adjust the gain/loss value based on frequency.
To specify external gain/loss corrections:
1.
Press the
Ampl
button to display the
Amplitude control panel.
2.
Select the External Gain/Loss Correction
tab.
3.
To apply a constant correction factor,
select the
External Gain value
check
box. Enter a value into the text entry box.
NOTE.
You can enter either positive or
negative gain values. Positive values
represent a gain. Negative values represent
a loss.
4.
To use a table that corrects for gain/loss
values depending on frequency, click a
check box under
External Loss Tables
.
You can enable all three external loss
tables at the same time. Thus, you could
use different external loss tables for an
antenna, a preampli
fi
er, and for cabling.
5.
To edit the contents of an external loss
table, click the
Edit
button for the table
you want to edit.
RSA6100A Series Quick Start User Manual
53
Summary of Contents for RSA6120A series
Page 3: ......
Page 7: ......
Page 9: ...Table of Contents ii RSA6100A Series Quick Start User Manual...
Page 17: ...Preface x RSA6100A Series Quick Start User Manual...
Page 29: ...Installation 12 RSA6100A Series Quick Start User Manual...
Page 77: ...Operation 60 RSA6100A Series Quick Start User Manual...