Installation
Inspecting the Instrument
Launch the diagnostics application (Tools > Diagnostics). If failures occur, perform the following steps to get more information
about them. You can also use the following steps as a detailed incoming inspection to verify the functionality of your
instrument. If you want to check the accuracy speci
fi
cations of your instrument, see the
RSA6100A Series Real-Time
Spectrum Analyzers Speci
fi
cations and Performance Veri
fi
cation Technical Reference
manual PDF (Tektronix part number
077-0251-XX) located on the Documents CD.
1.
Select
Tools > Diagnostics
.
2.
On the Diagnostics tab, click
All
Modules, All Tests
.
3.
Click
Single Test
.
4.
Click
RUN
.
The instrument will run through
the tests one at a time. A check
mark or X icon will appear to the
right of each check box as that
test is completed. The check icon
means the test was passed. The
X means the test was failed.
5.
For tests that require manual
intervention, follow the on-screen
instructions to complete the tests.
NOTE.
Select the Diagnostics Failure Information tab to see basic diagnostic failure information. Use the Windows Event
Viewer (a subset of the Administrative tools in the Windows Control Panel) to view failure history and nondiagnostic failures
reported by the application.
10
RSA6100A Series Quick Start User Manual
Summary of Contents for RSA6120A series
Page 3: ......
Page 7: ......
Page 9: ...Table of Contents ii RSA6100A Series Quick Start User Manual...
Page 17: ...Preface x RSA6100A Series Quick Start User Manual...
Page 29: ...Installation 12 RSA6100A Series Quick Start User Manual...
Page 77: ...Operation 60 RSA6100A Series Quick Start User Manual...