![Tektronix RSA6120A series Quick Start User Manual Download Page 58](http://html1.mh-extra.com/html/tektronix/rsa6120a-series/rsa6120a-series_quick-start-user-manual_1078170058.webp)
Operation
Power Triggering.
Power triggering triggers the instrument on time-domain signal characteristics. The incoming data is
compared to a user-selected level in dBm. You can select the time-domain bandwidth and trigger on the rising or falling edge.
In addition to de
fi
ning the trigger event, you can de
fi
ne the following parameters by selecting the appropriate tab in
the Trigger control panel:
Position and Delay parameters that de
fi
ne where a trigger occurs within the acquired data record and whether to
delay the trigger for a set amount of time.
Whether a single event can trigger an entire sweep, or each segment of the sweep requires a new trigger event in
swept acquisition mode.
RSA6100A Series Quick Start User Manual
41
Summary of Contents for RSA6120A series
Page 3: ......
Page 7: ......
Page 9: ...Table of Contents ii RSA6100A Series Quick Start User Manual...
Page 17: ...Preface x RSA6100A Series Quick Start User Manual...
Page 29: ...Installation 12 RSA6100A Series Quick Start User Manual...
Page 77: ...Operation 60 RSA6100A Series Quick Start User Manual...