Operating Basics
Changing Network
Settings
This procedure assumes that you have already installed the RFM220
instrument on your network and have powered the instrument on. For
fi
rst-time operation, perform the
fi
rst-time operation procedure. (See page 19,
Perform the following steps to con
fi
gure the network settings of a RFM220
instrument. You should always work with your network administrator to
correctly set these values.
CAUTION.
If you change the network settings of a RFM220 instrument, you
must reset the instrument before you use the Aggregator to start monitoring
the instrument.
1.
On the PC or server where the RFM220 Aggregator is installed, locate the
RFM220 Device Setup utility that was installed in the following directory:
Windows 7 systems:
C:\Program Files (x86)\Tektronix\RFM220\RFM220
Aggregator\RFM220DeviceSetup.exe
Windows XP systems:
C:\Program Files\Tektronix\RFM220\RFM220
Aggregator\RFM220DeviceSetup.exe
NOTE.
If you installed the RFM200 Aggregator software to a location other
than the default location, your path to the RFM220 Device Setup utility will be
different than the path shown above.
2.
Start the RFM220 Device Setup utility as follows for the operating system
on your PC or server:
Windows 7 systems. Right-click the
RFM220DeviceSetup.exe
fi
le
and select
Run as Administrator
.
Windows XP systems. Double-click the
RFM220DeviceSetup.exe
fi
le.
3.
In the RFM220 Device Setup dialog, enter the current IP address of
the RFM220 instrument in the Device IP Address box, and then click
Connect
.
If you cannot locate the IP address of the instrument, you can recover
the address using an RS-232 command. (See page 77,
RFM220 User Manual
49