Specification
and
Performance Check—PS 503A
Performance
Check Procedure
1. Check 0-20 Volt Output Voltage
a. Check that all
controls on the PS 503A front panel
are fully ccw and that the
400 mA MAX indicator light is off
(this
light
should
remain off as long as the PS 503A is
installed
in
the high-power
compartment of the power
module). Also check that both CURRENT LIMIT indicator
lights
are
off. (For instruments SN B022064-below, these
lights may
be on; if so, adjust the appropriate CURRENT
LIMIT control
cw just enough to turn them off).
b.
Set
the digital
voltmeter to the 2 volts dc range. Use
patch
cords
to connect
the +20 V floating
output
ter
minals of
the PS 503A to the digital voltmeter input
terminals;
connect
the red patch cord
from the 0—20V red
binding post
to the
digital
voltmeter high input terminal,
and
the black patch cord from the
COMMON (black)
binding
post to the
digital
voltmeter low input terminal.
c. Check
that the PS 503A + VOLTS controls are fully
ccw
and
check
for a meter reading
of .0000, within
±100 mV.
d. Set
the
digital voltmeter to the
200 volts dc range
and
set the PS 503A + VOLTS controls fully cw.
e.
Check
—for a
meter reading
of at least +20.1 volts,
but less than
+20.4,volts.
f.
Adjust
the PS 503A + VOLTS controls fully ccw and
move
the red patch cord from the 0—20V red binding post
to
the 0—20V green binding post on the
PS 503A.
g.
Repeat
parts
1c through 1e,
using the PS 503A
— VOLTS controls to check the —
20 V supply.
b.
Disconnect
the
red
patch cord from the digital
voltmeter
high input terminal and set the
digital voltmeter
to the
2000
mA de range. Reconnect the red patch cord to
the
digital
voltmeter high input terminal.
c.
Check—
for
a current reading of less than 100 mA
for the +20 V supply. The PS 503A
+ CURRENT LIMIT
light should
be
on.
d.
Move
the red patch
cord from the PS 503A 0—20V
red
binding post to the PS 503A
0—20V green binding
post.
e. Check—for a current reading of
less than 100 mA
for
the -20 V supply. The
PS 503A
- CURRENT LIMIT
light should
be on.
f.
Rotate the PS 503A
- CURRENT LIMITcontrol fully
cw.
g.
Check—
for
a current reading of more than
1000 mA, but less than
1500
mA for the —20 V supply.
h. Reconnect the red patch cord to the PS 503A
0—20V
red
binding
post and rotate the PS 503A +
CURRENT
LIMIT control
fully cw.
i.
Check
—for a current reading of more than
1000 mA,
but
less than 1500 mA for the +20 V supply.
j. Set all controls on the PS
503A front panel fully ccw
and
retain
the digital voltmeter connections to the
PS
503A.
h.
Set
the PS 503A —
VOLTS controls fully ccw and
move
the red patch cord back to the 0-20V red binding
post
on
the
PS 503A. Check to
be certain that all PS 503A
front-panel controls
are fully
ccw.
2.
Check
Variable
Current Limiting
a.
Adjust
the PS 503A
+ VOLTS controls for +5.0 V
output.
Move the red patch cord to the PS 503A 0—20V
green
binding post
and adjust
the PS 503A — VOLTS
controls for
—5.0 V output.
Move the red patch cord back
to
the
PS 503A 0—20V red binding post.
3. Check Load
Regulation
a. Disconnect
the red patch
cord from the digital
voltmeter
input terminal and set
the
digital voltmeter to the
20 volts
dc
range. Reconnect the red patch cord to the
digital voltmeter high
input terminal; the meter reading
should
indicate approximately zero volts. Both CURRENT
LIMIT lights on
the PS 503A front panel should
be off and
remain
off during the load regulation check.
b.
Rotate
the CURRENT LIMIT
control on the PS 503A
fully
cw.
2-6
Summary of Contents for PS 503A
Page 1: ...Tektronix COMMITTED TO EXCELLENCE PS 503A DUAL POWER SUPPLY INSTRUCTION MANUAL...
Page 4: ...Fig 1 1 PS 503A Dual Power Supply plug in module PS 503A...
Page 36: ...Options PS 503A OPTIONS No options are available at this time...
Page 41: ...I8S4 OX R V P FEB IR77 PS 5O3A...
Page 42: ...is 4 qx FL V P FEB IR77 Poweb Supply J bm bowo t above...
Page 45: ...TO...
Page 46: ...1834 04...
Page 47: ...PARTIAL POWER SUPPLY...
Page 52: ...REV A DEC 1975 21...
Page 53: ......