Specification and
Performance Check—PS 503A
Table 2-4 (cont)
Description
Performance
Requirements
Application
Example
Adapter
Bnc female to dual-banana
plugs.
Provides signal inter
connection.
Tektronix
Part No.
103-0090-00.
Load
resistor
20 Ω,
20 W, 5%,
(two
10 Ω, 10 W in series).
Load
tests.
Tektronix
Part No.
308-0246-00.
Load
resistor
5 Ω
,
5 W, 5%.
Load tests.
Tektronix
Part No.
308-0179-00.
Resistor
(2
required)
4.99 kΩ, 1/2 W,
1%
or
better.
Dual
tracking
detector
test.
Tektronix Part
No.
323-0260-00.
1X
passive probe
Compatible with
5A-series
amplifiers used
in
oscilloscope.
Provides signal inter
connection.
Tektronix
P6028
Probe.
'Requires
TM 500-Series Power Module.
Preliminary Procedure
1.
Ensure
that all test
equipment and the PS 503A
under test are
suitably adapted to the line voltage to
be
applied.
Refer
to the installation section of the power
module manual.
2. Be certain that
the POWER switch on
the power
module is
off and that the OUTPUT pushbutton on the
PS
503A is
in the out position.
3.
Install the PS 503A in the far right (high
power)
compartment
of
the power module.
4.
Connect
the
power module to the autotransformer
and
connect the autotransformer to the line voltage
source.
Set
the autotransformer
voltage output control for
zero
volt
output.
5. Set all front-panel controls on the PS 503A
fully
counterclockwise (ccw).
Apply power to the power
module
and press
in the OUTPUT pushbutton on the
PS
503A.
Push the PS 503A
VOLTS DUAL TRACKING
control
in.
6.
Adjust
the
autotransformer to the nominal line
voltage and check that the two VOLTS
indicator lights on
the PS 503A
are on.
7.
Set
the vertical deflection factor for 1 mV/div and
sweep
rate for 5 ms/div on the test oscilloscope. Use ac
coupling
for the
vertical
amplifier unit during this
procedure.
8. Connect all test equipment to a suitable line voltage
source.
9. Turn on all test equipment and allow at least twenty
minutes
for the equipment
to
warm-up and stabilize.
NOTE
During warm-up time,
attach banana pins
to
the
leads
of the
20
Ω
and
5
Ω
load
resistors so that they
can be
inserted in the holes
of the dual-banana
adapter.
This allows
the load resistors to be con
nected
as close
as
possible to the output terminals
on
the PS 503A so
contact resistance is kept to a
minimum.
2-5
Summary of Contents for PS 503A
Page 1: ...Tektronix COMMITTED TO EXCELLENCE PS 503A DUAL POWER SUPPLY INSTRUCTION MANUAL...
Page 4: ...Fig 1 1 PS 503A Dual Power Supply plug in module PS 503A...
Page 36: ...Options PS 503A OPTIONS No options are available at this time...
Page 41: ...I8S4 OX R V P FEB IR77 PS 5O3A...
Page 42: ...is 4 qx FL V P FEB IR77 Poweb Supply J bm bowo t above...
Page 45: ...TO...
Page 46: ...1834 04...
Page 47: ...PARTIAL POWER SUPPLY...
Page 52: ...REV A DEC 1975 21...
Page 53: ......