Specification
and
Performance Check—PS 503A
+5 V
Ground
Reference Supply
Table 2-1
(cont)
Characteristics
Performance
Requirements
Supplemental
Information
Load
Regulation
Within
3 mV with a 1
A load change.
Within 1
mV with
a 400 mA load
change.
Transient
Recover
≤20 µs for a constant voltage to
recover
within
20 mV for a
nominal
output voltage after a
400
mA change in output current.
Line
Regulation
Within
5 mV for a ±10% line
voltage change.
Ripple and
Noise
(2 Hz
to 2
MHz instrument
not
in current limiting.)
3 mV peak-to-peak or less with a
1
A load.
1
mV
peak-to-peak or less with
a 400 mA load
(noise and
ripple
increase
unpredictably
when current is being
limited).
Temperature
Coefficent
0.025%/°
C or less.
Stability
0.1%
+5
mV (or less) of drift in
8
hours.
Line voltage, load
and tem
perature
held constant.
Supply
Isolation from
Ground
350
V
(dc
+ peak ac).
Dual
Tracking Mode
Offset
Error
If the two
supplies are set indepen
dently
to any
given
voltage ratio, and
if both
supplies
are then
monitored
and
varied a given amount (by using
the
VOLTS
DUAL
TRACKING
control),
the two supplies will
maintain the
same voltage
ratio as initially set
within
50
mV; until one supply reaches
either
0
v' or 20 V.
Output
4.75
V dc to 5.25 V
dc with
a 5 ohm
load.
Measured at
a temperature be
tween
+20° C and +30° C.
Load
Regulation
Within
100 mV with a 1 A load change.
Line
Regulation
Within
50 mV for a ±10% line volt
age
change.
Measured
at a temperature
be
tween
+20° C and +30° C.
Ripple and Noise
(2
Hz to 2
MHz)
5
mV peak-to-peak or less with a 1 A
load.
Long Term
Stability
0.5% or less of
drift.
Overload
Protection
Fixed
(automatic) current limiting
and
over-temperature shutdown.
2-2
REV.
A,
AUG.
1977
i
Summary of Contents for PS 503A
Page 1: ...Tektronix COMMITTED TO EXCELLENCE PS 503A DUAL POWER SUPPLY INSTRUCTION MANUAL...
Page 4: ...Fig 1 1 PS 503A Dual Power Supply plug in module PS 503A...
Page 36: ...Options PS 503A OPTIONS No options are available at this time...
Page 41: ...I8S4 OX R V P FEB IR77 PS 5O3A...
Page 42: ...is 4 qx FL V P FEB IR77 Poweb Supply J bm bowo t above...
Page 45: ...TO...
Page 46: ...1834 04...
Page 47: ...PARTIAL POWER SUPPLY...
Page 52: ...REV A DEC 1975 21...
Page 53: ......