Functional Check and Calibration
Test Procedure
6.
Set the probe Input Mode to A–B.
7.
Adjust the oscilloscope to trigger on and
display a stable waveform (or press the
Autoset button).
NOTE.
If you do not see a waveform, check
the connection at the probe body. (See page 4,
Connect TriMode Adapters to the Probe Body.)
8.
When you see a stable square wave,
check the amplitude. (Use the horizontal
cursors.) Signal output levels for some
oscilloscope models are listed below.
DPO73304DX: 440 mV p-p
DPO72504DX: 440 mV p-p
P7600 Series TriMode Probes Quick Start User Manual
13
Summary of Contents for P7625
Page 1: ...x P7600 Series TriMode Probes ZZZ Quick Start User Manual P071302601 071 3026 01...
Page 2: ......
Page 3: ...P7600 Series TriMode Probes ZZZ Quick Start User Manual xx www tektronix com 071 3026 01...
Page 6: ......
Page 14: ...Preface viii P7600 Series TriMode Probes Quick Start User Manual...
Page 86: ...Maintenance 72 P7600 Series TriMode Probes Quick Start User Manual...