Section 2: Performance verification
DMM6500 6½-Digit Multimeter with Scanning Calibration and Adjustment Manual
2-42
DMM6500-905-01 Rev. E June 2021
2. Set the decade capacitor box to
0 F
.
3. On the DMM6500, press the
FUNCTION
key, select the
Measure Functions
tab, and select
Capacitance
.
4. Press the
MENU
key.
5. Under Measure, select
Settings
.
6. Set the Range to
1 nF
.
7. Press the
MENU
key.
8. Select
Calculations
and select
Rel
, then
Acquire
.
Cabling could be as high as ≈ 300 pF, which could prevent full
-scale verification due to the large
cable capacitance offset. Cable lengths should be minimized to reduce cable capacitance as much
as possible.