Section 3: Adjustment
DMM6500 6½-Digit Multimeter with Scanning Calibration and Adjustment Manual
3-30
DMM6500-905-01 Rev. E June 2021
Command parameters for an ac current adjustment
When calibrating your DMM6500 for an ac current adjustment, use the following command
parameters.
Send each command parameter twice. First, send the parameter using the
setup
command. Second,
send the parameter using the
execute
command.
The example code below uses the first line of code from the table below. Send all command
parameters using this command.
cal.adjust.step.setup("cal_ACI_100uA_1kHz_tenth")
cal.adjust.step.execute("cal_ACI_100uA_1kHz_tenth")
Command parameters
Calibrator
function
Calibrator
range
Calibrator
value
Calibrator
frequency
cal_ACI_100uA_1kHz_tenth
ac current 0.00022
0.00001
1 kHz
cal_ACI_100uA_1kHz_fs
ac current 0.00022
0.0001
1 kHz
cal_ACI_1mA_1kHz_tenth
ac current 0.00022
0.0001
1 kHz
cal_ACI_1mA_1kHz_fs
ac current 0.0022
0.001
1 kHz
cal_ACI_10mA_1kHz_tenth
ac current 0.0022
0.001
1 kHz
cal_ACI_10mA_10Hz_fs
ac current 0.022
0.01
10 Hz
cal_ACI_10mA_10kHz_fs
ac current 0.022
0.01
10 kHz
cal_ACI_10mA_1kHz_fs
ac current 0.022
0.01
1 kHz
cal_ACI_100mA_1kHz_tenth
ac current 0.022
0.01
1 kHz
cal_ACI_100mA_1kHz_fs
ac current 0.22
0.1
1 kHz
cal_ACI_100mA_10Hz_fs
ac current 0.22
0.1
10 Hz
cal_ACI_200mA_1kHz_fs
ac current 0.22
0.2
1 kHz
cal_ACI_200mA_10kHz_fs
ac current 0.22
0.2
10 kHz
cal_ACI_1A_1kHz_tenth
ac current 0.22
0.1
1 kHz
cal_ACI_1A_10Hz_fs
ac current 2.2
1
10 Hz
cal_ACI_1A_1kHz_fs
ac current 2.2
1
1 kHz
cal_ACI_3A_1kHz_tenth
ac current 2.2
0.3
1 kHz
cal_ACI_3A_1kHz_fs
ac current 2.2
2
1 kHz