DMM6500 6½-Digit Multimeter with Scanning Calibration and Adjustment Manual
Section 3: Adjustment
DMM6500-905-01 Rev. E June 2021
3-27
Prepare your DMM6500 for an ac voltage adjustment
To prepare to run ac adjustments:
1. Set the TERMINALS switch to
FRONT
.
2. Connect a cable between the calibrator and the DMM6500 as shown in the figure below.
Figure 27: Connection for an ac voltage adjustment
3. Allow the instrument and cables to settle for 30 seconds.
4. On the calibrator, source
10 mV 1.0 kHz
and allow the calibrator and DMM6500 to settle
properly.
5. Enable the
OPR
key.
6. Ensure that the
OPERATE
display is illuminated.
7. Allow the calibrator and cable to settle properly.