DMM6500 6½-Digit Multimeter with Scanning Calibration and Adjustment Manual
Section 3: Adjustment
DMM6500-905-01 Rev. E June 2021
3-31
AC current 10 A range adjustment
The following procedure provides instructions for completing a front-terminal adjustment ac current in
the 10 A range. For this adjustment, you need a Fluke 5720 calibrator. The following section provides
a command parameter table to complete the adjustment.
For this adjustment, you need a:
•
Fluke 5720 calibrator
•
10 A cable
Prepare your DMM6500 for an ac current 10 A range adjustment
To prepare the DMM6500 for an ac current adjustment in the 10 A range:
Connect the DMM6500 to the calibrator as shown in the following figure using 10 A cables.
Figure 29: Connection for a 10 A range adjustment