AC/DC Current Probe 
Model SL206 

USER MANUAL 

 

 

 

I

1 V/A

10 mV/A

OFF

ZERO

 

 

 

 

 

 

www.

GlobalTestSupply

.com

Quality AEMC Products Online at:

[email protected]

Summary of Contents for SL206

Page 1: ...AC DC Current Probe Model SL206 USER MANUAL I 1 V A 1 0 m V A O F F Z E R O www GlobalTestSupply com Quality AEMC Products Online at sales GlobalTestSupply com...

Page 2: ...ning the instrument to our repair and calibration facility for a nominal charge The recommended calibration interval for this instrument is 12 months and begins on the date of receipt by the customer...

Page 3: ...to us for repair provided we have your warranty registration information on file or a proof of purchase AEMC Instruments will at its option repair or replace the faulty material YOU CAN NOW REGISTER...

Page 4: ...rol and Connector Identification 5 Specifications 6 Electrical Specifications 6 Mechanical Specifications 7 Safety Specifications 8 Frequency Response Curves Typical 9 Operation 10 Connection 10 Zeroi...

Page 5: ...conductors which may create noise Check the magnetic mating surfaces of the probe jaws these should be free of dirt rust or other foreign matter Do not use a probe that is cracked damaged or has defec...

Page 6: ...ters Receiving Your Shipment Upon receiving your shipment make sure that the contents are consistent with the packing list Notify your distributor of any missing items If the equipment appears to be d...

Page 7: ...it Hall sensor technology senses the magnetic field produced by the current measured and generates a millivolt DC or AC output signal The narrow jaw is designed for optimum use in crowded wiring in in...

Page 8: ...0 to 100kHz 18mV DC to 1Hz 15 V 1Hz to 10kHz 140 V 10 to 100kHz 180 V 1 The maximum permitted value is only limited by the cross section of the primary conductor Saturation of the magnetic core of the...

Page 9: ...condensation Maximum Cable Diameter 0 46 11 8mm Zero Adjustment 20 turn potentiometer 1 5A min Handle Lexan 920 A UL 94 V2 Case Protection IP20 per IEC 529 Drop Test 1 0 m on 38mm of oak on concrete t...

Page 10: ...lution 2 Double insulation or reinforced insulation between primary or secondary and outer case of handle per IEC 1010 2 32 Electromagnetic Compatibility EN 50081 1 Class B EN 50082 2 Electrostatic di...

Page 11: ...rent Probe Model SL206 9 Frequency Response Curves Typical 0 10 20 30 40 50 60 1 10 100 1000 10000 Hz 1 1 2 2 1 mV mA 10 mV A www GlobalTestSupply com Quality AEMC Products Online at sales GlobalTestS...

Page 12: ...of residual magnetism often remains in the core This residual magnetism will cause the voltmeter to show a small DC reading even though there is no current passing through the jaws of the probe The re...

Page 13: ...cal shock do not attempt to perform any servicing unless you are qualified to do so Avoid electrical shock and or damage to the instrument do not get water or other foreign agents into the electronic...

Page 14: ...rded calibration data Chauvin Arnoux Inc d b a AEMC Instruments 15 Faraday Drive Dover NH 03820 USA Tel 800 945 2362 Ext 360 603 749 6434 Ext 360 Fax 603 742 2346 or 603 749 6309 repair aemc com Or co...

Page 15: ...www GlobalTestSupply com Quality AEMC Products Online at sales GlobalTestSupply com...

Page 16: ...Chauvin Arnoux Inc d b a AEMC Instruments 15 Faraday Drive Dover NH 03820 www aemc com 99 MAN 100029 v12 07 11 www GlobalTestSupply com Quality AEMC Products Online at sales GlobalTestSupply com...