Event log descriptions
Column title
Description
Example
Sequence
Each instrument maintains independent
sequence counters:
▪
One for each combination of UDP multicast
network interface and UDP multicast
destination port
▪
One for each TCP connection
Domain
Displays the LXI domain number; the default value
is 0 (zero)
0
Flags
Contain data about the LXI trigger packet;
values are:
▪
1 - Error
▪
2 - Retransmission
▪
4 - Hardware
▪
8 - Acknowledgments
▪
16 - Stateless bit
16
Data
The values for this are always 0 (zero)
Accessing the event log from the command interface
You can access the event log from any remote command interface. The event log must be enabled
before LXI trigger events can be viewed. To enable the event log, send:
eventlog.enable = 1
To view the event log from a remote interface, send:
print(eventlog.all())
This command outputs one or more strings similar to the following:
14:14:02.000 17 Jun 2019, LAN0, 10.80.64.191, LXI, 0, 1560780842, not available, 0,
0x10,0x00