Section 2: Installation
Series 2600B System SourceMeter® instrument User's Manual
2-4
2600BS-900-01 Rev. A / August 2021
Line frequency configuration
The factory configures the 2600B to detect the power line frequency automatically at each power-up.
This detected line frequency (either 50 Hz or 60 Hz) is used for aperture (NPLC) calculations.
In noisy environments, you can manually configure the instrument to match the actual line frequency.
To configure the line frequency from the front panel:
1. Press the
MENU
key, then turn the navigation wheel to select
LINE-FREQ
,
and then press the
ENTER
key.
2. Turn the navigation wheel
to select the appropriate frequency and then press the
ENTER
key. To
configure the instrument to automatically detect line frequency at each power-up, select
AUTO
.
3. Press the
EXIT (LOCAL)
key to return to the main display.
To configure the line frequency from a remote interface:
Set the
localnode.linefreq
or the
localnode.autolinefreq
attribute. To set the line
frequency to 60 Hz, send:
localnode.linefreq = 60
To configure the instrument to automatically detect line frequency at each power-up:
localnode.autolinefreq = true
System information
You can retrieve serial number, firmware revision, calibration dates, and memory usage
from the instrument.
To view the system information from the front panel:
1. Press the
MENU
key.
2. Select
SYSTEM
-
INFO
.
3. Select one of the following:
▪
FIRMWARE
▪
SERIAL#
▪
CAL
▪
MEMORY-USAGE
To retrieve system information from a remote interface:
To retrieve the firmware revision and serial number, send the
*IDN?
query.
For more information on system information, see the
Series 2600B Reference Manual
. In particular:
•
To determine memory usage, see the
meminfo()
function.
•
To determine when calibration was last run, see
smuX.cal.date
.
•
To determine when calibration is due, see
smuX.cal.due
.