Series 2268 850 Watt DC Power Supplies Reference Manual
Remote Operation
2268S-901-01 Rev. A / January 2015
157
Another benefit to the readback command is it allows the auto sequence program to be
captured and stored on the client side for reloading when the program is to persist beyond
power cycles.
Command:
[:]PROGram[<channel>]:READback?
For Example:
:PROG:START
*CLS
:VOLT 5.4
:CURR 0.25
*IDN?
:PROG:STOP
:PROG:READ?
Response:
*CLS
:VOLT 5.4
:CURR 0.25
*IDN?
:PROG:STOP
Reloading an Auto Sequence Program
The following procedure will demonstrate how to store a command beyond a power cycle of
the unit.
Important:
This functionality is only available through USB, RS-232 and
RS-485. It is not possible to reload an auto sequence program over the GPIB and
Ethernet/LAN interfaces.
To reload Auto Sequence program from file:
1.
Set the data rate of the device to be 1200 bps.
2.
Close the connection to the hyperterminal program and configure the connection
data rate to be 1200 bps.
3.
Execute the program recording command:
:PROG:START
Summary of Contents for Keithley 2268 Series