Appendix A: Interfaces
A-56
K1297-G20 Monitor User Manual
Defining CPU Numbers
The following section describes how to define CPU numbers (VME bus
addresses) if you work with a second Ethernet board consisting of a
Power-PC board plus hooked up mezzanines. Only an experienced service
technician should perform this setting procedure.
CAUTION.
The following steps must be performed at an ESD approved
workplace. Electrostatic discharge can permanently destroy components
that have to be temporarily removed from your device.
In order to operate several Ethernet boards within one device, the default
CPU numbers of the second (third, fourth) board have to be changed.
1. Check that the address is set correctly for each board. Typical addresses
are as follows:
Table A–32: Ethernet boards: address settings
Board
Address
1st Ethernet board
07 (with 08 implicitly); no changing necessary
2nd Ethernet board
05 (with 06 implicitly)
3rd Ethernet board
03 (with 04 implicitly)
NOTE.
Only select odd numbers when changing the address for an
additional Ethernet board, and follow the recommended settings as listed
in the table above.
The CPU numbers being stored within NVRAM of the Ethernet board are
an integral part of the CPU’s bootline. You can modify this bootline as
described in the following.
Summary of Contents for K1297-G20
Page 4: ......
Page 12: ...Service Safety Summary viii K1297 G20 Monitor User Manual...
Page 100: ...Examples 88 K1297 G20 Monitor User Manual...
Page 122: ...Appendix A Interfaces A 22 K1297 G20 Monitor User Manual Figure A 1 AP 4 AP 4 256 Board...
Page 139: ...Appendix A Interfaces K1297 G20 Monitor User Manual A 39 Figure A 5 V X board...
Page 158: ...Appendix A Interfaces A 58 K1297 G20 Monitor User Manual...
Page 170: ...Appendix B Keyboard B 12 K1297 G20 Monitor User Manual...
Page 190: ...Appendix D Windows XPe Troubleshooting E 12 K1297 G20 Monitor User Manual...