Table B- 1: Supported measurements and their definition (Cont.)
Name
Definition
Low
The value used as 0% whenever High Ref, Mid Ref, and Low Ref values are needed (as in fall
time and rise time measurements). May be calculated using either the min/max or the histogram
method. With the min/max method it is the minimum value found. With the histogram method, it
refers to the most common value found below the midpoint. Measured over the entire waveform
or gated region.
Maximum
Voltage measurement. The maximum amplitude. Typically the most positive peak voltage.
Measured over the entire waveform or gated region.
Mean
Voltage measurement. The arithmetic mean over the entire waveform or gated region.
Minimum
Voltage measurement. The minimum amplitude. Typically the most negative peak voltage.
Measured over the entire waveform or gated region.
Negative Duty Cycle
Timing measurement of the first cycle in the waveform or gated region. The ratio of the negative
pulse width to the signal period expressed as a percentage.
NegativeDutyCycle
=
NegativeWidth
Period
×
100%
Negative Overshoot
Voltage measurement. Measured over the entire waveform or gated region.
NegativeOvershoot
=
Low–Min
Amplitude
×
100%
Negative Width
Timing measurement of the first pulse in the waveform or gated region. The distance (time)
between MidRef (default 50%) amplitude points of a negative pulse.
Peak to Peak
Voltage measurement. The absolute difference between the maximum and minimum amplitude
in the entire waveform or gated region.
Phase
Timing measurement. The amount one waveform leads or lags another in time. Expressed in
degrees, where 360
_
comprise one waveform cycle.
Period
Timing measurement. Time it takes for the first complete signal cycle to happen in the waveform
or gated region. The reciprocal of frequency. Measured in seconds.
Positive Duty Cycle
Timing measurement of the first cycle in the waveform or gated region. The ratio of the positive
pulse width to the signal period expressed as a percentage.
PositiveDutyCycle
=
PositiveWidth
Period
×
100%
Summary of Contents for CSA7404B
Page 4: ......
Page 16: ...Table of Contents xii CSA7000B Series TDS7000B Series Instruments User Manual ...
Page 24: ...Product Description 1 4 CSA7000B Series TDS7000B Series Instruments User Manual ...
Page 130: ...Acquiring Waveforms 3 62 CSA7000B Series TDS7000B Series Instruments User Manual ...
Page 186: ...Triggering 3 118 CSA7000B Series TDS7000B Series Instruments User Manual ...
Page 234: ...Measuring Waveforms 3 166 CSA7000B Series TDS7000B Series Instruments User Manual ...
Page 324: ...Data Input Output 3 256 CSA7000B Series TDS7000B Series Instruments User Manual ...
Page 330: ...MyScope 3 262 CSA7000B Series TDS7000B Series Instruments User Manual ...
Page 364: ...Appendix A Specifications A 30 CSA7000B Series TDS7000B Series Instruments User Manual ...
Page 414: ...Glossary Glossary 18 CSA7000B Series TDS7000B Series Instruments User Manual ...
Page 438: ...Index Index 24 CSA7000B Series TDS7000B Series Instruments User Manual ...