Creating and Using Math Waveforms
3- 190
CSA7000B Series & TDS7000B Series Instruments User Manual
To turn on a real or imaginary spectrum, touch the Math button, the
Define/Edit Expression Editor button, and then select the Freq tab. Touch
either the Real or the Imag menu items to enter an expression. Then touch
the Ch tab and one of the channel buttons. Touch apply.
Using the Phase Controls.
You can set the vertical units to degrees, radians, or
seconds of group delay. You select these choices by touching the Math button,
the Spectral Analysis Setup button, and then selecting the Phase tab. Select the
desired scale type from Degrees, Radians, or Group Delay.
H
Phase Reference Position. Phase is a relative measurement that must have a
time domain reference point. The phase value is specified with respect to this
phase reference position.
For the spectral analyzer, the phase reference position is the 50% position of
the gate, that is, the middle of the gate interval of the data that is input to the
spectral analyzer. This is true for all window functions except for the Tek
Exponential window. This window has the reference point at the 20%
position of the gate.
H
Phase Unwrap. The spectral analyzer produces phase values from --
π
to
π
radians or --180 to 180 degrees. However, when you perform impulse
response testing and the phase is continuous, then phase values outside these
ranges may occur. The spectral analyzer then wraps the data with disconti-
nuities in the display from +180 to --180 degrees. Phase unwrap will display
the correct result by unwrapping the phase.
Phase unwrap is only valid when the phase spectrum is a continuous
function of frequency. Therefore, do not use it when analyzing the harmonic
content of the typical repetitive signal.
H
Suppression Threshold. Random noise in the spectrum may have phase
values over the entire range. This could make the phase display unusable.
However, you can set the suppression threshold control to a level in dB. The
phase of any complex spectral points with a magnitude below this threshold
is set to zero.
H
Phase Unwrap Algorithm. The algorithm searches for the largest magnitude
in the current span. Phase unwrap is then performed in both directions in
frequency from that point. This results in a stable phase unwrap.
H
Phase Spectrum dejitter. An instrument acquisition system jitters by one
sample interval. Signals at the Nyquist frequency only have two samples per
cycle. This would cause 180
_
of phase jitter if it were not corrected. The
instrument phase spectrum is dejittered so that accurate measurements of
phase are obtained from DC to the Nyquist frequency.
Summary of Contents for CSA7404B
Page 4: ......
Page 16: ...Table of Contents xii CSA7000B Series TDS7000B Series Instruments User Manual ...
Page 24: ...Product Description 1 4 CSA7000B Series TDS7000B Series Instruments User Manual ...
Page 130: ...Acquiring Waveforms 3 62 CSA7000B Series TDS7000B Series Instruments User Manual ...
Page 186: ...Triggering 3 118 CSA7000B Series TDS7000B Series Instruments User Manual ...
Page 234: ...Measuring Waveforms 3 166 CSA7000B Series TDS7000B Series Instruments User Manual ...
Page 324: ...Data Input Output 3 256 CSA7000B Series TDS7000B Series Instruments User Manual ...
Page 330: ...MyScope 3 262 CSA7000B Series TDS7000B Series Instruments User Manual ...
Page 364: ...Appendix A Specifications A 30 CSA7000B Series TDS7000B Series Instruments User Manual ...
Page 414: ...Glossary Glossary 18 CSA7000B Series TDS7000B Series Instruments User Manual ...
Page 438: ...Index Index 24 CSA7000B Series TDS7000B Series Instruments User Manual ...