Measuring Waveforms
3- 144
CSA7000B Series & TDS7000B Series Instruments User Manual
Reference level calculation methods
High (50 mV)
High reference
Mid reference (0 mV)
Low reference
Low (--50 mV)
90%
10%
--40 mV
40 mV
50%
0 mV
Figure 3- 37: Reference-level calculation methods
The High and Low levels from which the reference levels are calculated are the
levels established using the selected Hi/Low method described on page 3--142.
Use the procedure that follows to quickly take a measurement based on the
default settings for High/Low and for reference-levels.
Overview
To take automatic measurements
Related control elements and resources
Prerequisites
1.
Obtain a stable display of the waveform to be measured.
See page 3--29 for acquisition setup and page 3--63 for
trigger setup.
To Take Automatic
Measurements
Summary of Contents for CSA7404B
Page 4: ......
Page 16: ...Table of Contents xii CSA7000B Series TDS7000B Series Instruments User Manual ...
Page 24: ...Product Description 1 4 CSA7000B Series TDS7000B Series Instruments User Manual ...
Page 130: ...Acquiring Waveforms 3 62 CSA7000B Series TDS7000B Series Instruments User Manual ...
Page 186: ...Triggering 3 118 CSA7000B Series TDS7000B Series Instruments User Manual ...
Page 234: ...Measuring Waveforms 3 166 CSA7000B Series TDS7000B Series Instruments User Manual ...
Page 324: ...Data Input Output 3 256 CSA7000B Series TDS7000B Series Instruments User Manual ...
Page 330: ...MyScope 3 262 CSA7000B Series TDS7000B Series Instruments User Manual ...
Page 364: ...Appendix A Specifications A 30 CSA7000B Series TDS7000B Series Instruments User Manual ...
Page 414: ...Glossary Glossary 18 CSA7000B Series TDS7000B Series Instruments User Manual ...
Page 438: ...Index Index 24 CSA7000B Series TDS7000B Series Instruments User Manual ...