background image

80A09 26 GHz ESD Protection Device Instructions

BNC T connector.

Two BNC cables (example: Tektronix part number 012-0076-xx, 20 inch
BNC cable).

Two BNC to Banana plug adapters.

Precautions while

performing the test

This test checks the 80A09 leakage and clamping thresholds. Use caution when
operating the Source Measure Unit (SMU) in this test setup. Even though the
80A09 protects the 80E04 from reasonable overvoltages, use care to avoid
damaging the 80A09. Follow these precautions:

Always disconnect the 80A09 from the sampling module input for this test.

Always set SMU test limits as speci

fi

ed.

Always disable the SMU output before connecting or disconnecting the SMU
supply from the test setup.

Set the SMU at zero volts before enabling the output.

Procedure

CAUTION.

Follow ESD handling guidlelines when connecting cables to the

80A09. (See page 1, Best practices to prevent ESD.)

1.

If connected, remove the cable from the 80A09.

2.

Remove the 80A09 from the 80E04 module input.

CAUTION.

Do not perform this test with the 80A09 connected to the 80E04. Doing

so can damage the 80E04 module input.

Do not connect a ground or 50

terminator to the 80A09.

3.

Con

fi

gure the SMU before connecting to the 80A09:

Set the SMU at zero volts.

Disable the SMU supply output.

4.

Connect the adapters and cables as shown in the following diagram.

80A09 Instructions

3

Summary of Contents for 80A09

Page 1: ... following guidelines to prevent ESD when using the 80A09 or anytime that you are connecting signals or handling the 80A09 or the 80E04 Electrical Sampling Module Always wear a grounded antistatic wrist strap when handling or connecting the 80A09 to the 80E04 module when connecting cables to the 80A09 or when handling or installing sampling modules in the mainframe instrument Wear anti static clot...

Page 2: ...5 in lb 2 Connect the cable from the DUT to the 80A09 Make sure to discharge the cable center and shield to ground before connecting to the 80A09 Use a torque wrench to tighten the connector to 7 5 in lb CAUTION To prevent damaging the 80A09 80E04 module connectors do not put excess pull or lateral movement on the cable attached to the 80A09 80E04 assembly Verify that 80A09 is functional Use the f...

Page 3: ...test limits as specified Always disable the SMU output before connecting or disconnecting the SMU supply from the test setup Set the SMU at zero volts before enabling the output Procedure CAUTION Follow ESD handling guidlelines when connecting cables to the 80A09 See page 1 Best practices to prevent ESD 1 If connected remove the cable from the 80A09 2 Remove the 80A09 from the 80E04 module input C...

Page 4: ...utput 2 0 VDC Verify that the measured current is less than 100 μA 8 Set the SMU to source 65 mA with the overvoltage limit set to 3 5 V Verify that the measured voltage is 3 42 VDC 9 Set the SMU to sink 65 mA with the overvoltage limit set to 3 5 V Verify that the measured voltage is 3 42 VDC 10 Set the SMU output to zero volts disable the SMU output and disconnect the SMA to BNC adapter from the...

Page 5: ...ropriately This symbol indicates that this product complies with the applicable European Union requirements according to Directives 2002 96 EC and 2006 66 EC on waste electrical and electronic equipment WEEE and batteries For information about recycling options check the Support Service section of the Tektronix Web site www tektronix com Restriction of hazardous substances This product is classifi...

Reviews: