background image

80A09 26 GHz ESD Protection Device Instructions

NOTE.

Make sure to connect the ground (GND ) side of each BNC to Banana plug

adapter to the LO connectors of the SMU.

5.

Use the following guidelines while using the SMU with the 80A09:

Do not exceed the test voltage value.

Once tested, reduce the voltage to zero volts before changing SMU

settings.

Do not disconnect or reconnect any test cables until you reduce the output

to zero.

6.

Set the SMU to output

+2.0 VDC

. Verify that the measured current is less

than

+100

μ

A

.

7.

Set the SMU to output

–2.0 VDC

. Verify that the measured current is less

than

–100

μ

A

.

8.

Set the SMU to source

+65 mA

, with the overvoltage limit set to

+3.5 V

.

Verify that the measured voltage is

3.42 VDC

.

9.

Set the SMU to sink

–65 mA

, with the overvoltage limit set to

–3.5 V

. Verify

that the measured voltage is

–3.42 VDC

.

10.

Set the SMU output to

zero

volts, disable the SMU output, and disconnect the

SMA to BNC adapter from the 80A09.

CAUTION.

If any of the checks are outside the stated limits, the 80A09 device

can no longer provide ESD protection. Do not use a defective 80A09 with an
80E04. Contact Tektronix Customer Support to order a replacement 80A09 ESD
Protection Device.

11.

Repeat all of these steps to test other 80A09 devices.

4

80A09 Instructions

Summary of Contents for 80A09

Page 1: ... following guidelines to prevent ESD when using the 80A09 or anytime that you are connecting signals or handling the 80A09 or the 80E04 Electrical Sampling Module Always wear a grounded antistatic wrist strap when handling or connecting the 80A09 to the 80E04 module when connecting cables to the 80A09 or when handling or installing sampling modules in the mainframe instrument Wear anti static clot...

Page 2: ...5 in lb 2 Connect the cable from the DUT to the 80A09 Make sure to discharge the cable center and shield to ground before connecting to the 80A09 Use a torque wrench to tighten the connector to 7 5 in lb CAUTION To prevent damaging the 80A09 80E04 module connectors do not put excess pull or lateral movement on the cable attached to the 80A09 80E04 assembly Verify that 80A09 is functional Use the f...

Page 3: ...test limits as specified Always disable the SMU output before connecting or disconnecting the SMU supply from the test setup Set the SMU at zero volts before enabling the output Procedure CAUTION Follow ESD handling guidlelines when connecting cables to the 80A09 See page 1 Best practices to prevent ESD 1 If connected remove the cable from the 80A09 2 Remove the 80A09 from the 80E04 module input C...

Page 4: ...utput 2 0 VDC Verify that the measured current is less than 100 μA 8 Set the SMU to source 65 mA with the overvoltage limit set to 3 5 V Verify that the measured voltage is 3 42 VDC 9 Set the SMU to sink 65 mA with the overvoltage limit set to 3 5 V Verify that the measured voltage is 3 42 VDC 10 Set the SMU output to zero volts disable the SMU output and disconnect the SMA to BNC adapter from the...

Page 5: ...ropriately This symbol indicates that this product complies with the applicable European Union requirements according to Directives 2002 96 EC and 2006 66 EC on waste electrical and electronic equipment WEEE and batteries For information about recycling options check the Support Service section of the Tektronix Web site www tektronix com Restriction of hazardous substances This product is classifi...