80A09 26 GHz ESD Protection Device Instructions
Install the 80A09
You insert the 80A09 devices between the 80E04 TDR sampling module inputs
and the cables to the device under test, with the 80A09 connected directly to
the 80E04.
CAUTION.
Follow ESD handling guidlelines when connecting the 80A09 to the
module and when connecting cables to the 80A09. (See page 1, Best practices
to prevent ESD.)
1.
Connect the 80A09 to the 80E04 module. Use a torque wrench to tighten the
connector to 7.5 in-lb.
2.
Connect the cable from the DUT to the 80A09. Make sure to discharge the
cable center and shield to ground before connecting to the 80A09. Use a
torque wrench to tighten the connector to 7.5 in-lb.
CAUTION.
To prevent damaging the 80A09/80E04 module connectors, do not
put excess pull or lateral movement on the cable attached to the 80A09/80E04
assembly.
Verify that 80A09 is functional
Use the following procedure to verify that the 80A09 is fully functional to protect
the sampling module from ESD (do these steps before major testing sessions, or
as part of a scheduled maintenance routine):
Required test equipment
Source Measure Unit (SMU).
Adapter, K to BNC.
2
80A09 Instructions