background image

80A09 26 GHz ESD Protection Device Instructions

Install the 80A09

You insert the 80A09 devices between the 80E04 TDR sampling module inputs
and the cables to the device under test, with the 80A09 connected directly to
the 80E04.

CAUTION.

Follow ESD handling guidlelines when connecting the 80A09 to the

module and when connecting cables to the 80A09. (See page 1, Best practices
to prevent ESD.)

1.

Connect the 80A09 to the 80E04 module. Use a torque wrench to tighten the
connector to 7.5 in-lb.

2.

Connect the cable from the DUT to the 80A09. Make sure to discharge the
cable center and shield to ground before connecting to the 80A09. Use a
torque wrench to tighten the connector to 7.5 in-lb.

CAUTION.

To prevent damaging the 80A09/80E04 module connectors, do not

put excess pull or lateral movement on the cable attached to the 80A09/80E04
assembly.

Verify that 80A09 is functional

Use the following procedure to verify that the 80A09 is fully functional to protect
the sampling module from ESD (do these steps before major testing sessions, or
as part of a scheduled maintenance routine):

Required test equipment

Source Measure Unit (SMU).

Adapter, K to BNC.

2

80A09 Instructions

Summary of Contents for 80A09

Page 1: ... following guidelines to prevent ESD when using the 80A09 or anytime that you are connecting signals or handling the 80A09 or the 80E04 Electrical Sampling Module Always wear a grounded antistatic wrist strap when handling or connecting the 80A09 to the 80E04 module when connecting cables to the 80A09 or when handling or installing sampling modules in the mainframe instrument Wear anti static clot...

Page 2: ...5 in lb 2 Connect the cable from the DUT to the 80A09 Make sure to discharge the cable center and shield to ground before connecting to the 80A09 Use a torque wrench to tighten the connector to 7 5 in lb CAUTION To prevent damaging the 80A09 80E04 module connectors do not put excess pull or lateral movement on the cable attached to the 80A09 80E04 assembly Verify that 80A09 is functional Use the f...

Page 3: ...test limits as specified Always disable the SMU output before connecting or disconnecting the SMU supply from the test setup Set the SMU at zero volts before enabling the output Procedure CAUTION Follow ESD handling guidlelines when connecting cables to the 80A09 See page 1 Best practices to prevent ESD 1 If connected remove the cable from the 80A09 2 Remove the 80A09 from the 80E04 module input C...

Page 4: ...utput 2 0 VDC Verify that the measured current is less than 100 μA 8 Set the SMU to source 65 mA with the overvoltage limit set to 3 5 V Verify that the measured voltage is 3 42 VDC 9 Set the SMU to sink 65 mA with the overvoltage limit set to 3 5 V Verify that the measured voltage is 3 42 VDC 10 Set the SMU output to zero volts disable the SMU output and disconnect the SMA to BNC adapter from the...

Page 5: ...ropriately This symbol indicates that this product complies with the applicable European Union requirements according to Directives 2002 96 EC and 2006 66 EC on waste electrical and electronic equipment WEEE and batteries For information about recycling options check the Support Service section of the Tektronix Web site www tektronix com Restriction of hazardous substances This product is classifi...

Reviews: