Removal and installation
Figure 17: Sample of electrical module hardware removal
Replacing the chassis
CAUTION.
Transport the sampling module chassis with terminations on the inputs.
Transport the sampling module chassis in the original container and store the
sampling modules on a metal or conductive surface.
When handling the exposed chassis sampling module only, do not touch any
exposed component on the circuit board. Hold the exposed module by the edges of
the circuit board or the rear or front housings.
1.
Reinstall the chassis into the cover by reversing the removal procedure. (See
page 54,
WARNING.
Do not install or remove any sampling modules while the instrument
is powered on. Electrical shock may occur. Always power the instrument down
before attempting to remove or insert any sampling module to avoid potential
injury from shock.
2.
Install the electrical sampling module into one of the instrument small module
compartments.