Removal and installation
Removal and installation
WARNING.
Before performing any procedure in this section, disconnect the power
cord from the line voltage source. Failure to do so could cause serious injury
or death.
Only Quali
fi
ed personnel should perform service procedures. Before performing
this procedure, read the General and Service Safety summaries at the beginning
of this manual.
CAUTION.
Perform all DSA8300 removal and installation procedures in a
static-controlled environment, following the precautions listed in the Preventing
ESD section. (See page 19.)
NOTE.
Read the cleaning procedure before disassembling the instrument for
cleaning.
If you replace any circuit board, check to see if it is necessary to adjust the
instrument. (See Table 4 on page 68.)
Required equipment
Most assemblies in this instrument can be removed with a screwdriver handle
mounted with a size T-15, Torx® screwdriver tip. Use this tool whenever a
procedure step instructs you to remove or install a screw unless a different size
screwdriver is speci
fi
ed in that step.
Table 1: Required equipment for removal and replacement
Item No.
Name
Description
Part number
1
Screwdriver handle
Accepts Torx®-driver bits
General Tool: 620-440
2
Screwdriver handle, extended
length, 10.5 to 12 inches
Accepts Torx®-driver bits and 3/16 socket tip. Used
for removing the T-10 screws and 3/16 inch hex posts
from the electrical and optical module slots.
Standard tool or Xcelite
99X10 V N/S#48706
3
T-10 Torx tip
Torx®-driver bit for T-10 size screw heads, used for
removing the electrical or optical module chassis.
General Tool: 640-235
4
T-15 Torx tip
Used for removing most the screws. Torx®-driver bit
for T-15 size screw heads.
General Tool: 640-247
5
T-20 Torx tip
Used for removing the handle hardware. Torx®-driver
bit for T-20 size screw heads.
General Tool: 640-250
6
3/16 inch socket tip
Used for removing the 3/16 inch hex posts from the
electrical and optical module slots.
Standard tool
7
1/8 inch
fl
at-bladed screwdriver
Screwdriver for unlocking cable connectors.
Standard tool
24
DSA8300 Digital Serial Analyzer and Modules Service Manual