Calibration
—
Type
543B/RM543B
MAGNIFIED
position and
reposition the vertical trace to
the
exact
horizontal center
of
the graticule. (The HORIZONTAL
POSITION
controls will appear extermely sensitive to
touch
in
the X100 position.) Now turn the HORIZONTAL DIS
PLAY switch
back to the NORMAL (X1) position and adjust
SWP MAG
REGIS R359 so that the trace falls exactly under
the
cente.r graticule line. Repeat this
adjustment several
times
to
obtain
exact coincidence of the
trace for
both
positions
of the HORIZONTAL DISPLAY switch. Recheck
step
13
before continuing to step 18.
Set
the
HORIZONTAL DISPLAY switch to NORMAL (X1),
the
triggering controls for a stable display and the INTEN
SITY control for normal
brilliancy. Position
the start of the
display to
the left side
of the
graticule. Observe the marker
that
appears
near the center of the graticule and position
the
display so that the center
marker is
aligned with the
center
graticule
line. Then
turn the
HORIZONTAL DISPLAY
switch
to the X100 SWEEP MAGNIFIED
position and posi
tion
the marker near the center
graticule line so that ihe
leading
edge is exactly aligned with the center graticule
line. Turn
the HORIZONTAL DISPLAY switch through all
SWEEP
MAGNIFIED positions
back to NORMAL (X1) and
observe
the
leading edge of the marker;
it should remain
stationary,
aligned with the center line of the graticule.
18. Check
Sweep
Rates
—
1 msec
to
5
sec
With
the
HORIZONTAL
DISPLAY
switch set to NORMAL
(XI)
and the triggering control set for a stable display,
check
time base sweep rates according to the following
table:
TIME/CM
Switch
Setting
Time
Markers
Markers
Displayed
.1
mSEC
100 μsec
1 /cm
.2
mSEC
100
μsec
2/cm
.5 mSEC
500
μsec
1
/
cm
1
mSEC
1 msec
1
/
cm
2
mSEC
1
msec
2/cm
5 mSEC
5
msec
1
/cm
10 mSEC
10
msec
1
/cm
20
mSEC
10
msec
2/cm
50
mSEC
50
msec
1
/cm
.1
SEC
100 msec
1/cm
.2
SEC
100
msec
2/cm
.5
SEC
500 msec
1 /cm
1
SEC
1
sec
1/cm
2 SEC
1
sec
2/cm
5
SEC
5
sec
1
/cm
19.
Check
Variable
Time/Cm
Control
and
Uncali
brated
Neon.
The VARIABLE control provides
for a
complete range of
control
between
the calibrated TIME/CM steps. To check
the
operation of this control, set TIME/CM to 1 mSEC
CALIBRATED. Connect
5 msec markers to the vertical input
connec'or
and
trigger
the oscilloscope for c stable display
consisting of 1 marker/5
cm. Next,
turn
the VARIABLE
control
fully counterciockwise. The display should now con
sist
of markets every 2 cm or less. Check to see that the
UNCALIBRATED
neon indicator
lamp is lit in all positions
of
the VARIABLE control except
when switched to the CALI
BRATED position.
Set
the front-panel
controls as shown in Table 5-3 before
proceeding
with step
20.
TABLE
5-3
Crt
Controls
INTENSITY
FOCUS
ASTIGMATISM
SCALE
ILLUM
Usable level
As is
As
is
As is
Time
Base
TRIGGERING
LEVEL
STABILITY
TRIGGERING
MODE
TRIGGER SLOPE
VARIABLE
(TIME/CM)
TIME/CM
HORIZONTAL
DISPLAY
SINGLE
SWEEP
0
Clockwise
AC
-PINT
CALIBRATED
.1
mSEC
NORMAL (X1)
NORMAL
SWEEP
Other
Controls
HORIZONTAL
POSITION
Midrange
VERNIER
(HORIZONTAL
POSITION)
EXTERNAL HORIZONTAL
VOLTS/CM
Midrange
10
VARIABLE (EXTERNAL
HORIZONTAL VOLTS/
CM) CALIBRATED
AMPLITUDE
CALIBRATOR
TU-7
Verfical
Position
Test
Function
Variable
Other Controls
OFF
Centered
Low
Load
Adjust
for desired
amplitude
As
is
20. Adjust
Time
Base
Sweep
Rates
50
μsec
/cm
to
0.02
μsec
/cm.
Apply 10 µsec markers from the Type 180A
to the vertical
input
connector,
and adjust the triggering controls for a
stable
display.
Turn
the HORIZONTAL DISPLAY switch to
X5 SWEEP
.MAGNIFIED and horizontally position the trace
so tha
t
the first time
marker
is aligned with the center
graticule
line. Then
switch the TIME/CM switch to 50 μSEC
and
check for horizontal
shift of the first marker. If
shift
occurs,
adjust C337 (see Fig. 5-3) until
the first marker of
both
the .1 μSEC
and 50 µSEC positions occur at the same
point.
Turn the HORIZONTAL
DISPLAY switch to NORMAL
(X
1
), Time Base A TIME/CM
to 10 μSEC and proceed with
the
following
adjustments:
5-7
Summary of Contents for 543B
Page 4: ...The Type 543B Oscilloscope...
Page 38: ...Parts List Type 543B FRONT REAR 6 2...
Page 46: ...Parts List Type 543B CHASSIS 6 10...
Page 54: ...Parts List Type 543B CABLE HARNESS CERAMIC STRIP DETAIL 6 18...
Page 75: ...CMD fe 4 TYPE 54 SB OSCILLOSCOPE BLOCK DIAGRAM...
Page 77: ...A 3b5 POWER SUPPLY O...
Page 79: ...CMD CRT CJRCUIT 4 TYPL543B OSCILL05COPL A...
Page 81: ...SEE PARTS LIST FOR SEMICONDUCTOR TYPES DON 9fo5 TYPE 5zl3B OSCILLOSCOPE VERTICAL AMPLIFIER L...
Page 85: ......
Page 88: ...TYPE S43B OSCILLOSCOPE c HORIZONTAL AMPLIFIES 7...