Calibration
—
Type
543B/RM543B
Fig. 5-2.
Location of internal
adjustments and test points
for time base, power supply, and amplitude calibrator.
To
make this adjustment, connect the voltmeter between
the
CAL
TEST PT
jack and ground (see Fig. 5-2),
and adjust
CAL
ADJ R879 for a reading of e100 volts. To
assure
suitable symmetry
of the calibrator waveform, the
reading
at
this point should not be less than 45 nor more
than
55 volts
when the calibrator is turned on. Readings
outside this
range
are generally caused by unbalanced
multivibrator tubes
(V875 or V885A).
3.
High-Voltage
Power
Supply
Adjustment
Connect
the voltmeter
between ground and the high-
voltage
check
point (sec Fig. 5-3),
and set the HIGH VOLT
AGE
control
(see Fig. 5-3) for a meter
reading of exactly
—
1700
volts. Disconnect the voltmeter.
Check
the regulation of the high-voltage
power supply
by
turning
the INTENSITY control to 7 and
defocusing the
trace
with
the FOCUS and ASTIGMATISM controls.
Place
the HORIZONTAL
DISPLAY
switch to EXT and position the
defocused
spot to
the left
side of
the crt. Now observe the
spot while
adjusting the line voltage 10% above and below
the design center voltage.
The spot should
not have shown
any
"blooming".
4.
Trace
Alignment
Position the HORIZONTAL
DISPLAY switch
to NORMAL
(X
1
)
and adjust the INTENSITY, FOCUS, and ASTIGMA
TISM
controls
to obtain a focused trace of normal intensity.
Position the
trace behind
the center
horizontal grati
cule
line.
If the trace
and graticule line do not coincide
over
the width
of the graticule, adjust the
TRACE ROTA
TION
control until they
do.
5.
Geometry
The
geometry of the crt display is adjusted by means of
the
GEOM control.
To
achieve optimum linearity, verti
cal
lines
are displayed on the crt and the GEOM control
is
adjusted
for minimum curvature of
the lines. Nonlinearity
is
most
noticeable at the edges of the graticule.
Connect
500
μsec
markers from the Type 180A to the Ext
Input
connector of
the Type TU-7 and position the base line
of
the markers below the bottom
of the crt face so it is not
visible.
Adjust
the Type
TU-7 Variable control so that
the
markers
over-scan the crt.
Obtain a stable display with the
triggering
controls,
and adjust the GEOM control (see Fig.
5-3)
for straight vertical lines
running parallel to the left
and
right
edges of the graticule (see Fig. 5-4).
NOTE
The
amplitude
calibrator may be used for
this
step,
but
due to the
low intensity of the vertical
lines, the
adjustment is somewhat more difficult.
6.
Vertical
Amplifier
Low-Frequency
Adjustments
Set the
Test Function switch of the Type
TU-7 to Common
Mode and
adjust DC BAL
R495
until the trace is super
imposed
on
the center graticule
line.
5-3
Summary of Contents for 543B
Page 4: ...The Type 543B Oscilloscope...
Page 38: ...Parts List Type 543B FRONT REAR 6 2...
Page 46: ...Parts List Type 543B CHASSIS 6 10...
Page 54: ...Parts List Type 543B CABLE HARNESS CERAMIC STRIP DETAIL 6 18...
Page 75: ...CMD fe 4 TYPE 54 SB OSCILLOSCOPE BLOCK DIAGRAM...
Page 77: ...A 3b5 POWER SUPPLY O...
Page 79: ...CMD CRT CJRCUIT 4 TYPL543B OSCILL05COPL A...
Page 81: ...SEE PARTS LIST FOR SEMICONDUCTOR TYPES DON 9fo5 TYPE 5zl3B OSCILLOSCOPE VERTICAL AMPLIFIER L...
Page 85: ......
Page 88: ...TYPE S43B OSCILLOSCOPE c HORIZONTAL AMPLIFIES 7...