Calibration
—
Type
543B/RM543B
Other
Controls
HORIZONTAL
POSITION
VERNIER
(HORIZONTAL
POSITION)
EXTERNAL
HORIZONTAL
VOLTS/CM
VARIABLE
(EXTERNAL
HORIZONTAL
VOLTS/
CM)
AMPLITUDE CALIBRATOR
Midrange
Midrange
10
CALIBRATED
OFF
Before
installing the Type TU-7
and applying power to
the instrument
the resistances of the power supplies should
be
checked. The typical resistances of the supplies may
be
found in
the chart below.
Nominal
Resistances
of
Power
Supplies
Supply
Approx. Resistance
to
Ground
-150
2k
+ 100
2k
+225
2k
+350
1
k
+500
15k
Fig.
5-1. Low-voltage
power
supply fest point locations.
Install
the Type TU-7 and preset its controls as follows:
Vertical
Position
Centered
Test
Function
Low
Load
Other Controls
As
is
Connect
the power cord and
the ac voltmeter to the
out
put
of
the
autotransformer. Turn
the power switch to the
ON
position and adjust the autotransformer
for an output
voltage
to match
the design center
voltage of the
oscillo
scope.
Allow the instrument to warm up for several minutes
before
proceeding with the
calibration adjustments. During
calibration,
periodically
check the input voltage to the
instrument
and
adjust the autotransformer as
necessary to
maintain
the
voltage
at the design center
voltage except
when
the
power supply regulation
is being checked.
CAUTION
Do not reset
the —150
control
unless the power
supply
voltages are actually out
of
tolerance
or
you
are
planning to perform a complete calibra
tion
of the
instrument.
Check
the
delay time delay
relay. The relay armature
should pull
in with a "click" sound after 15 to 60 seconds
has
elapsed.
PROCEDURE
1.
Low-Voltage
Power
Supplies
Measure the
output voltage of the
—150v, +100 v,
+
225
v, +350 v,
and +500 v regulated supplies at the
points indicated,
in
Fig.
5-1. The output voltage of the
—150-volt
and
the
other regulated supplies must be within
3%
of
their rated values. Set the —150 control (see Fig.
5-
2) so
that all
of these
voltages are within the specified
tolerance.
To
check the
regulation of the power supplies, set
the
Test
Function switch
of the Type TU-7 to High Load and
adjust
the line voltage for a voltage 10%
lower than your
design center
voltage. Now check the voltage of each
supply.
The —150v, +100v, +225 v
and
+350 v supplies
should
still be
within 3% of their proper values. The 4-500 v
supply
should still be within 5% of its proper value.
The
power
supply
ripple is checked by connecting a one
times
probe
from the test oscilloscope to the supply being
checked.
The table below gives the
approximate ripple
amplitudes
of each power
supply.
After the
power supplies have been checked on low
line
voltage,
the line voltage should be raised 10% above
design
center
voltage and the Test
Function switch of the
Type
TU-7 set to Low
Load. Repeat the voltage and ripple
checks. The
same limits apply.
When
the power supply regulation checks are
completed
return
the
line
voltage to the design center voltage.
Typical
Ripple
Amplitudes
Supply
Typical Ripple
-150
5
mv
+
100
10 mv
+
225
5 mv
+350
20
mv
+
500
20 mv
2.
AMPLITUDE
CALIBRATOR
Adjustments
CAL
ADJ
R879
should be set to provide exactly 4-100
volts
at
the CAL TEST PT when the AMPLITUDE CALIBRATOR
switch
is in rhe
OFF position. Under these conditions, the
CAL
OUT voltages should
be
within 3% of the front-panel
readings.
5-2
Summary of Contents for 543B
Page 4: ...The Type 543B Oscilloscope...
Page 38: ...Parts List Type 543B FRONT REAR 6 2...
Page 46: ...Parts List Type 543B CHASSIS 6 10...
Page 54: ...Parts List Type 543B CABLE HARNESS CERAMIC STRIP DETAIL 6 18...
Page 75: ...CMD fe 4 TYPE 54 SB OSCILLOSCOPE BLOCK DIAGRAM...
Page 77: ...A 3b5 POWER SUPPLY O...
Page 79: ...CMD CRT CJRCUIT 4 TYPL543B OSCILL05COPL A...
Page 81: ...SEE PARTS LIST FOR SEMICONDUCTOR TYPES DON 9fo5 TYPE 5zl3B OSCILLOSCOPE VERTICAL AMPLIFIER L...
Page 85: ......
Page 88: ...TYPE S43B OSCILLOSCOPE c HORIZONTAL AMPLIFIES 7...