Calibration
— Type
543B/RM543B
voltage 10%
above
and below the design center line volt
age. From
the stable trace position at low-line voltage to
a
possible new
stable position at high-line
voltage, the trace
should
not drive more than 0.5 cm.
7.
Checking
Alternate
Trace
and
Chopped
Blanking
Set
the
AMPLITUDE CALIBRATOR to
20 VOLTS and the
Type
TU-7
Test Function
switch to Alternate. Connect a
jumper from
the CAL OUT to
the Ext Input connector of the
TU-7. Center
the display with
the Vertical Position control
and
check each sweep rate of the TIME/CM switch for
the
alternating
traces.
When
observing slow
sweep rates, the spot will consist
of
a
single dot when the upper trace crosses the crt. How
ever,
the
spot will consist of two
dots, one above the other,
when
the lower trace crosses the crt.
Set the
Type TU-7
Test Function switch to Chopped and
the
TIME/CM control to 5 μSEC. Adjust the triggering con
trols for a
stable display. Switch
the CRT CATHODE SELEC
TOR
to
CHOPPED
BLANKING and observe that the
vertical
lines
(chopped transients) disappear. Return the CRT CATH
ODE
SELECTOR switch to the
EXTERNAL CATHODE posi
tion
and reset
the
oscilloscope and Type TU-7 controls
according
to
Table 5-1.
TABLE
5-1
Crt
Controls
Usable
level
As is
As
is
As
is
Time
Base
INTENSITY
FOCUS
ASTIGMATISM
SCALE
ILLUM
TRIGGERING
LEVEL
STABILITY
TRIGGERING
MODE
TRIGGER
SLOPE
VARIABLE
(TIME/CM)
TIME/CM
HORIZONTAL DISPLAY
SINGLE
SWEEP
0
Clockwise
AUTO
+EXT
CALIBRATED
.5
mSEC
NORMAL (X
1)
NORMAL
SWEEP
Other
Controls
HORIZONTAL
POSITION
VERNIER
(HORIZONTAL
POSITION)
EXTERNAL HORIZONTAL
VOLTS/CM
VARIABLE
(EXTERNAL
HORIZONTAL VOLTS/
CM)
AMPLITUDE
CALIBRATOR
Midrange
Midrange
10
CALIBRATED
100
VOLTS
TU-7
Vertical
Position
Centered
Test
Function
Gain
Set
Other Controls
As
is
TRIGGERING
ADJUSTMENTS
8.
Trigger
Sensitivity
With
no input leads connected to the Type TU-7 and the
front-panel controls
set as
in
Table 5-1, connect a probe
from
the test oscilloscope to the junction of C131 and R49
and a
jumper from the TRIGGER INPUT connector to
ground.
Set
the test oscilloscope sweep rate for 10 msec/
cm, the vertical controls for ac coupling and input
signal of
5 volts
in amplitude, and the triggering controls for a stable
display.
Adjust the
TRIG SENS control until one cycle of the wave
form
occupies about 2.5 cm
of the sweep. Disconnect
the
jumper
between
the
TRIGGER INPUT connector and ground.
9.
Trigger
Level
Centering
Set
the TRIGGER SLOPE control
to +INT and the TRIG
GERING
MODE
switch to AC. Connect a test lead from the
CAL
OUT connector to the Ext Input connector on the Type
TU-7.
A signal with an amplitude of
4 cm should
be ob
served.
If
this signal amplitude
is not present recheck steps
2 and
6.
After observing the
4-cm signal, reduce rhe AMPLI
TUDE
CALIBRATOR signal
to 5 volts to obtain
a calibrated
2-mm
signal for use in the following procedure,
Center
the
trace
vertically on the crt and adjust the
INTENSITY,
FOCUS and ASTIGMATISM controls for best
definition.
Then ground the
junction of R19 and R20 with a
short
clip lead. This
junction is located
on top of the trigger
switch (see
Fig. 5-3).
Preset
the TRIG LEVEL CENTERING
control fully clock
wise.
Turn
the STABILITY control counterclockwise until the
trace
just disappears from the crt screen, then two or three
degrees further
counterclockwise.
Turn
the TRIG LEVEL CENTERING
control counterclock
wise
until the
display reappears on
the crt. Then switch the
TRIGGER SLOPE
control to —INT; it may be necessary to
turn
the TRIG LEVEL
CENTERING control clockwise slight
ly
to obtain a stable display. Then
while switching back and
forth between
+INT and —INT,
slightly readjust the TRIG
LEVEL
CENTERING control
for
sfab'e triggering in both
positions.
10.
Internal
Triggering
Dc
Level
Set
the Type TU-7 Test Function switch to Low
Load and
the
oscilloscope
AMPLITUDE
CALIBRATOR to 1
VOLT. Use
the
Type TU-7 Variable control to reduce the signal ampli
tude
to 6 mm.
Center
the
display vertically, and turn the TRIGGERING
MODE
switch to the DC position. While
switching the
TRIGGER SLOPE control
back and
forth between
+INT
and
—INT,
adjust
the INT TRIG DC LEVEL ADJ control for stable
triggering in
both positions. It may be necessary to slightly
readjust the TRIG
LEVEL CENTERING
control
to obtain
stable
triggering.
5-5
Summary of Contents for 543B
Page 4: ...The Type 543B Oscilloscope...
Page 38: ...Parts List Type 543B FRONT REAR 6 2...
Page 46: ...Parts List Type 543B CHASSIS 6 10...
Page 54: ...Parts List Type 543B CABLE HARNESS CERAMIC STRIP DETAIL 6 18...
Page 75: ...CMD fe 4 TYPE 54 SB OSCILLOSCOPE BLOCK DIAGRAM...
Page 77: ...A 3b5 POWER SUPPLY O...
Page 79: ...CMD CRT CJRCUIT 4 TYPL543B OSCILL05COPL A...
Page 81: ...SEE PARTS LIST FOR SEMICONDUCTOR TYPES DON 9fo5 TYPE 5zl3B OSCILLOSCOPE VERTICAL AMPLIFIER L...
Page 85: ......
Page 88: ...TYPE S43B OSCILLOSCOPE c HORIZONTAL AMPLIFIES 7...