Calibration
Procedure—
Type
535A
TRIG.
LEVEL CENTERING
1NT.
TRIG.
D.C.
LEVEL ADJ.
HIGH
TEST POINT
TRIG. SENS.
C364
C375
C33O
LOCKOUT
LEVEL
ADJ.
Fig.
6-3.
Top View of the oscilloscope showing locations of internal controls.
To
check
the
above
supplies for proper regulation, vary the
autotransformer voltage
between 105 v and
125 v
(or from
210
v to 250 v if the power
transformer is connected for
234
v
operation). All
of the regulated voltages should re
main
essentially constant.
The
ripple present on any
of the regulated supplies, as
measured with a test scope at
the voltage
check points, will
be
well under
10 mv., with AMPLITUDE CALIBRATOR-OFF,
and
neither sweep
operating.
2.
AMPLITUDE-CALIBRATOR
Adjustment
The Cal.
Adj. control should be set to provide a de out
put
of 100 volts when the AMPLITUDE CALIBRATOR switch
is
in the OFF
position. Under these conditions, the calibrator
output will
be
within 3% of the front-panel readings.
To
make this
adjustment connect the voltmeter between
the
Cai. Test
Pt. jack
and ground (see Right Side View,
Fig.
6-2), turn the AMPLITUDE CALIBRATOR switch to the
OFF
position, and adjust the Cal. Adj.
control for
a reading
of
exactly
100
volts. To assure suitable symmetry of the
calibrator
waveform,
the reading at this point should
not
be less than
45 v nor more than 55 volts when the cali
brator
is
turned on (to any of the output voltage settings).
Readings outside
this range are generally
caused by un
balanced multivibrator
tubes (V875 or V885A).
3.
High-Voltage
Power Supply Adjustment
This
adjustment determines the total accelerating poten
tial
on the crt, and thus affects the deflection sensitivity.
Connect
the voltmeter between ground and the high-
voltage
check point (see Top-View, Fig. 6-3), and set the H.
V.
Ad|. (see Right Side
View, Fig. 6-2) for a meter reading
of
exactly
—
1350 volts,
if your meter
has a full scale read
ing of
1200 v, you
can connect
the
positive meter lead to
the
—
1 50
v
supply, at any convenient point, and set the H.V.
Adj. for a
meter reading of —1200 v
4.
CRT Alignment
If
the crt has been
replaced, or if, due to
considerable
handling,
the trace does not align with the graticule, you
should
make this
adjustment before proceeding with the
calibration.
With no
vertical signal applied, free run
TIME BASE A, by
6-3
Summary of Contents for 535A
Page 2: ......
Page 4: ...Type 535A ...
Page 28: ...Circuit Description Type 535A 4 2 Fig 4 1 Simplified Type 535A Vertical Amplifier ...
Page 30: ...Circuit Description Type 535A 4 4 Fig 4 2 Simplified Time Base Trigger Circuit ...
Page 32: ...Circuit Description Type 535A Fig 4 3 Simplified Time Base Generator 4 6 ...
Page 34: ...Circuit Description Type 535A 4 8 Fig 4 4 Simplified Delay Pickoff Circuit ...
Page 38: ...Circuit Description Type 535A Fig 4 6 Simplified Alternate Trace Operation Circuitry 4 12 ...
Page 39: ...Circuit Description Type 535A 4 13 ...
Page 41: ...Circuit Description Type 535A Fig 4 8 Simplified Low Voltage Power Supply 4 15 ...
Page 42: ...Circuit Description Type 535A 4 16 Fig 4 9 Simplified Calibrator Circuit ...
Page 44: ...Circuit Description Type 535A Fig 4 10 Simplified Cathode Ray Tube Circuit 4 18 ...
Page 126: ...MRU ll IG 61 O Z ...
Page 138: ...C897 R896 R891 R890 R889 R888 R887 R886 R885 AMPLITUDE CALIBRATOR SWITCH Bottom View 535A A ...
Page 143: ...FIG 1 FRONT SWITCHES TVDC COCA ocril I iCrODC ...
Page 144: ...70 ...
Page 145: ...1 FIG 3 SWEEP CHASSIS HIGH VOLTAGE BOARD DELAY SWEEP CHASSIS ...
Page 146: ...1 FIG 4 POWER CHASSIS RECTIFIER MOUNTING PLATE A ...
Page 147: ...3 34 79 ...
Page 148: ...TYPE 535A OSCILLOSCOPE ...
Page 149: ......
Page 150: ...FIG 7 CABINET FRAME RAILS ...
Page 152: ...FIG 8 STANDARD ACCESSORIES 6 TYPE 535A OSCILLOSCOPE ...