General
The Type
535A Oscilloscope is a versatile
instrument which
is
adaptable to
a great number of applications. However,
to
make use of the
full potentialities of the instrument, it is
necessary that
you understand completely the operation of
each
control.
This
portion of the
Operator's Manual is in
tended
to
provide you with the basic information that you
require.
If
you
are familiar with other Tektronix Oscillo
scopes,
you should have very little difficulty in understanding
the
operation
of
the Type 535A since the function of many
controls
is the same as the function of corresponding con
trols on other
Tektronix instruments.
Plug-In Units
The Type
535A Oscilloscope is designed to
operate with
any
one of the
Tektronix letter-series plug-in units. The
particular
plug-in unit used must be selected by you to sat
isfy
the
requirements of your application. In selecting the
plug-in unit
for any particular application, you must con
sider
the
bandpass, sensitivity, and type
of input
required
for
that application. The plug- in units available will satisfy
the
requirements for most applications.
Preparation for Use
When
the
plug-in unit has been selected, insert it
into the
plug-in
compartment of the oscilloscope and
press firmly to
insure that the connectors
make proper
contact. Tighten
the
plug-in unit locking control to hold the unit securely in
place
and turn the oscilloscope INTENSITY control fully
counterclockwise.
Refer to
the voltage input marking on the back panel of
the
instrument to
verify that it is wired for the line voltage
you have available.
Refer
to
the maintenance section of this
manual
if
you need to change the power
transformer and
fan
wiring for a different input voltage.
Connect
the
power
cord to the rear of
the instrument and
to
the power line and place the POWER switch in the ON
position.
Cooling
A
fan maintains safe
operating temperature
by circulating
filtered
air over the rectifiers and other components. When in
operation, the
instrument must be placed so that
the air in
take
at the back
is clear of any obstruction that might im
pede
the
flow of air. Side panels should also be in place for
proper air circulation.
The air
filter should be kept clean,
in accordance with
cleaning instructions found in
the Main
tenance
Section of this manual.
Under
no circumstances should
your Oscilloscope be op
erated
without the fan running. Without the fan, inside tern-
ON 2
OPERATING
INSTRUCTIONS
perature
of the oscilloscope will rise to a dangerous level
in five
to ten minutes. In this event, the thermal cutout
switch
will
disconnect the power and keep it disconnected
until
the temperature drops
to a
safe level.
Time
Delay
Time
delay relays
used in the Type 535A Oscilloscope
delay
operation of
the instrument
for approximately 25 sec
onds
after
the
POWER switch is turned on to allow a brief
tube-warmup
period.
The delay allows the tubes sufficient
time
to heat
before the dc operating voltages are applied.
If
the ac power is off for only an instant, the normal 25-
second
delay will occur before the instrument returns to full
operation. This
delay will occur regardless of
whether the
ac
power is off because
of a momentary power failure or
is turned
off
with
the POWER switch.
Focus and Astigmatism
Controls
The
FOCUS
and ASTIGMATISM controls operate in con
junction
with each
other to allow you to obtain a sharp,
clearly
defined spot
or trace. The
proper setting of the
ASTIGMATISM
control is obtained by rotating the FOCUS
control
fully clockwise, setting the HORIZONTAL DISPLAY
switch to EXT.
×10, and adjusting the INTENSITY
control to
obtain
a spot
on the screen. After positioning the spot
to
the
center of the screen, the ASTIGMATISM control is ad
justed
for
the most nearly circular spot possible. The FOCUS
control
is then adjusted to reduce
the spot diameter as
much
as
possible.
(You
must be certain that
all input signals
to
the
oscilloscope are disconnected when you adjust the
FOCUS and
ASTIGMATISM
controls.)
Intensity Control
The
INTENSITY control is used to adjust the brightness
of
the oscilloscope display. This permits you to compen
sate
for changes
in brightness
resulting from changes in the
sweep
or triggering rate. The INTENSITY control is rotated
clockwise
to increase brightness and counterclockwise
to
decrease brightness. Care must
be taken when using the
INTENSITY
control that
the brightness is not turned up to
the point
where
the face of the cathode-ray tube is per
manently
damaged. If the intensity of the beam
is turned up
too far,
the
phosphor on the
face of the crt may be burned.
The
intensity of the beam should never be turned up to
the
point
where a halo
forms around the spot.
Graticule Illumination Control
The graticule
used with the Type 535A Oscilloscope is
accurately
marked with
10 horizontal
and 6 vertical 1-centi
meter divisions
with
2-millimeter markings on the center
lines.
These graticule markings allow
you to make time and
voltage
measurements from the oscilloscope screen.
2-1
Summary of Contents for 535A
Page 2: ......
Page 4: ...Type 535A ...
Page 28: ...Circuit Description Type 535A 4 2 Fig 4 1 Simplified Type 535A Vertical Amplifier ...
Page 30: ...Circuit Description Type 535A 4 4 Fig 4 2 Simplified Time Base Trigger Circuit ...
Page 32: ...Circuit Description Type 535A Fig 4 3 Simplified Time Base Generator 4 6 ...
Page 34: ...Circuit Description Type 535A 4 8 Fig 4 4 Simplified Delay Pickoff Circuit ...
Page 38: ...Circuit Description Type 535A Fig 4 6 Simplified Alternate Trace Operation Circuitry 4 12 ...
Page 39: ...Circuit Description Type 535A 4 13 ...
Page 41: ...Circuit Description Type 535A Fig 4 8 Simplified Low Voltage Power Supply 4 15 ...
Page 42: ...Circuit Description Type 535A 4 16 Fig 4 9 Simplified Calibrator Circuit ...
Page 44: ...Circuit Description Type 535A Fig 4 10 Simplified Cathode Ray Tube Circuit 4 18 ...
Page 126: ...MRU ll IG 61 O Z ...
Page 138: ...C897 R896 R891 R890 R889 R888 R887 R886 R885 AMPLITUDE CALIBRATOR SWITCH Bottom View 535A A ...
Page 143: ...FIG 1 FRONT SWITCHES TVDC COCA ocril I iCrODC ...
Page 144: ...70 ...
Page 145: ...1 FIG 3 SWEEP CHASSIS HIGH VOLTAGE BOARD DELAY SWEEP CHASSIS ...
Page 146: ...1 FIG 4 POWER CHASSIS RECTIFIER MOUNTING PLATE A ...
Page 147: ...3 34 79 ...
Page 148: ...TYPE 535A OSCILLOSCOPE ...
Page 149: ......
Page 150: ...FIG 7 CABINET FRAME RAILS ...
Page 152: ...FIG 8 STANDARD ACCESSORIES 6 TYPE 535A OSCILLOSCOPE ...