![Tektronix 2221A User Manual Download Page 95](http://html.mh-extra.com/html/tektronix/2221a/2221a_user-manual_1078835095.webp)
Measuring Signals
C l Step 2:
Using either probes or coaxial cables with equal time delays,
connect a known reference signal to the Channel 1 input and the un
known signal to the Channel 2 input.
f l Step 3:
Switch the Vertical Mode to
BOTH
and then select either
ALT
or
CHOP
H Step 4:
Set both
VOLTS/DIV
switches and both variable controls so the
displays are equal in amplitude.
f l Step 5:
Set the
SOURCE
to
CH 1
so the oscilloscope uses only the
reference signal for triggering. Adjust the
TRIGGER LEVEL
control for a
stable display.
f l Step 6:
Set the
SEC/DIV
switch to a sweep speed that displays about
one full cycle of the reference waveform.
f~ l
Step
7: Position the displays and adjust the
SEC/DIV
variable control so
that one cycie of the reference signal occupies exactly 8 horizontal
divisions. Each horizontal division of the graticule now represents 45° of
the cycle (360°
8 divisions).
F l Step 8:
Measure the horizontal difference in divisions between the
signals and multiply it by 45° (Figure 3-39).
Channel 1 Reference
Channel 2
Figure 3-39: Measuring Phase Difference
For higher resolution measurements of the phase difference the display may
be magnified horizontally by pulling the X I0 magnification knob out
(Figure 3-40).
2221A User Manual
3-39
Summary of Contents for 2221A
Page 1: ...User Manual Tektronix 2221A Digital Storage Oscilloscope 070 8156 02 ...
Page 6: ...It Welcome ...
Page 14: ...X 2221A User Manual ...
Page 16: ...xii Product Description ...
Page 26: ...Front Panel CRT Display Figure 2 2 Front Panel View Left Side 2 2 At a Glance ...
Page 48: ...2 24 At a Glance ...
Page 86: ...lopiayiiiij 3 30 In Detail ...
Page 100: ...Saving Waveforms and Setups 3 44 En Detail ...
Page 108: ...Appendix A Options and Accessories Appendices ...
Page 156: ...Appendix C Performance Verification j f i f P P j r fe t E F E J E J E fer 4 52 Appendices ...