![Tektronix 2221A User Manual Download Page 22](http://html.mh-extra.com/html/tektronix/2221a/2221a_user-manual_1078835022.webp)
Checking the Probe Compensation
NOTE
Always compensate a probe for the particular channel that you use
it with.
Use the following procedure to check the probe compensation:
Step 1
: Set the instrument controls as described in
Setting up the
Display.
n §tep 2: Connect the probe to the channel you intend to use it for.
{If the probe is properly “coded” the volts-per-division readout for the chan
nel will change to match the attenuation factor of the probe. The 10X probes
supplied with the oscilloscope already have the proper coding.)
□ SteP 3:
Clip the probe tip to the
PRB ADd
connector.
f~|
Step
4: Use the
VOLTS/DIV
control to display about 5 divisions of the
waveform,
□ Step
5: Use the
VERTICAL POSITION
and
HORIZONTAL POSITION
controls to center the display.
F I Step 6: Check the waveform against Figure 1-3 to see if the probe is
correctly compensated.
r
Over compensated
r
r
v .
Under compensated
Correctly compensated
Figure 1-3: Checking Probe Compensation
f~l Step 7: Adjust the compensation if necessary.
7-6
Before You Begin
Summary of Contents for 2221A
Page 1: ...User Manual Tektronix 2221A Digital Storage Oscilloscope 070 8156 02 ...
Page 6: ...It Welcome ...
Page 14: ...X 2221A User Manual ...
Page 16: ...xii Product Description ...
Page 26: ...Front Panel CRT Display Figure 2 2 Front Panel View Left Side 2 2 At a Glance ...
Page 48: ...2 24 At a Glance ...
Page 86: ...lopiayiiiij 3 30 In Detail ...
Page 100: ...Saving Waveforms and Setups 3 44 En Detail ...
Page 108: ...Appendix A Options and Accessories Appendices ...
Page 156: ...Appendix C Performance Verification j f i f P P j r fe t E F E J E J E fer 4 52 Appendices ...