Appendix C: Performance Verification
Procedure Steps:
Q Step 1: Check internal Triggering
a. Connect the leveled sine wave generator output via a 50 Q cable
and a 50 Q termination to the CH 1 OR X input connector,
b. Set the generator to produce a 10 MHz, 3.5 division display.
c. Set the
CH 1 VOLTS/DiV
switch to 50 mV.
d. CHECK — Stable display can be obtained by adjusting the Trigger
LEVEL
control for each switch combination given in Table A-12.
Table A-12: Switch Combinations for Triggering Checks
Trigger Mode
Trigger SLOPE
NORM
Positive
NORM
Negative
P-P AUTO
Negative
P-P AUTO
Positive
e. Set:
Vertical
MODE
CH 2
SOURCE
CH 2
f.
Move the cable from the CH 1 OR X input connector to the CH 2 OR
Y input connector.
g. Repeat part d.
h. Set:
SEC/DiV
0.1 ps
X I0 Magnifier
On (knob out)
i.
Set the generator to produce a 60 MHz, 1.0 division display.
j.
Repeat part d.
k. Set:
Vertical MODE
CH 1
SOURCE
CH 1
l.
Move the cable from the CH 2 OR Y input connector to the CH 1 OR
X input connector.
m. Repeat part d.
n. Set:
SEC/DiV
0.05 ps
o. Set the generator to produce a 100 MHz, 1.5 division display.
p. Repeat part d.
A-44
Appendices
Summary of Contents for 2221A
Page 1: ...User Manual Tektronix 2221A Digital Storage Oscilloscope 070 8156 02 ...
Page 6: ...It Welcome ...
Page 14: ...X 2221A User Manual ...
Page 16: ...xii Product Description ...
Page 26: ...Front Panel CRT Display Figure 2 2 Front Panel View Left Side 2 2 At a Glance ...
Page 48: ...2 24 At a Glance ...
Page 86: ...lopiayiiiij 3 30 In Detail ...
Page 100: ...Saving Waveforms and Setups 3 44 En Detail ...
Page 108: ...Appendix A Options and Accessories Appendices ...
Page 156: ...Appendix C Performance Verification j f i f P P j r fe t E F E J E J E fer 4 52 Appendices ...