Adjustment Procedure—2215A Service
g. CHECK— Display amplitude is 0.05 division or less.
h. Disconnect the test equipment from the instrument.
16. Check Com m on-Mode Rejection Ratio
a. Set:
VOLTS/DIV (both)
10 mV
INVERT
On (button in)
Channel 2 Input Coupling
DC
b.
Connect the leveled sine-wave generator output via a
50-fi cable, a 50-SJ termination, and a dual-input coupler to
the CH 1 OR X and CH 2 OR Y input connectors.
c.
Set the generator to produce a 25-MHz, 6-division
display.
d.
Vertically center the display using the Channel 2 POSI
TION control. Then set the VERTICAL MODE switch to
CH 1 and vertically center the display using the Channel 1
POSITION control.
e.
Set the VERTICAL MODE switches to BOTH and
ADD.
f. CHECK— Display amplitude is 0.6 division or less.
g.
If the check in part f meets the requirement, skip to
part p. If it does not, continue with part h.
h. Set the VERTICAL MODE switch to CH 1.
i.
Set the generator to produce a 50-kHz, 6-division
display.
j. Set the VERTICAL MODE switch to BOTH.
k.
Adjust the CH 1 or CH 2 VOLTS/DIV Variable control
for minimum display amplitude.
I. Set the VERTICAL MODE switch to CH 1.
m. Set the generator to produce a 25-MHz, 6-division
display.
n. Set the VERTICAL MODE switch to BOTH.
o. CHECK— Display amplitude is 0.6 division or less.
p. Disconnect the test equipment from the instrument.
17. Check Input G ate Current
a. Set:
VERTICAL MODE
CH 1
VOLTS/DIV (both)
2 mV
VOLTS/DIV Variable (both)
CAL detent
Input Coupling (both)
GND
b.
Position the trace on the center horizontal graticule
line using the Channel 1 POSITION control.
c.
CHECK— For 0.1 division or less trace shift while
alternating the Channel 1 Input Coupling switch between the
AC and GND positions.
d. Set the VERTICAL MODE switch to CH 2.
e. Repeat parts b and c using the Channel 2 controls.
NOTE
To continue with the Adjustm ent Procedure, remove
the instrum ent cabinet and allow a 20-minute time pe
riod to elapse before continuing with the Adjustm ent
Procedure. See the ‘Cabinet’ removal instructions lo
cated in the “Maintenance" section o f the manual.
5-12
Summary of Contents for 2215A
Page 1: ......
Page 9: ...2215A Service 4735 01 vi The 2215A Oscilloscope ...
Page 138: ...2215A Service 4735 04 Figure 9 4 Basic block diagram ...
Page 139: ......
Page 145: ...CIRCUIT BCARD N TER CONNECTIONS 2 2 1 5 A 4755 20 ...
Page 153: ...2215A S e rv ic e D UJ o C D GO CO in 10 CJ Figure 9 8 Circuit view of A1 Main board ...
Page 157: ...1 p 2 m 3 p 4 p 5 P 6 p 7 p 8 p 9 A T B Z 2 I5 A ...
Page 158: ...H T J t K t L t M T N T p t S 4 7JS Z2 ...
Page 168: ...I 2 3 4 5 6 7 8 9 B H K L M N 2 2 I 5 A 473S Z ...
Page 173: ...I 2 to 3 to 4 to 5 to 6 7 to 8 to 9 2 215A B TIMING ALTERNATE B SWEEP S ...
Page 176: ......
Page 181: ...I 2 3 4 5 6 7 8 9 AXIS CRT ...
Page 189: ...2215A Service 4735 11 ...
Page 193: ...3 ...
Page 194: ...2215A SERVICE ...
Page 195: ...4 7 5 ...
Page 200: ......
Page 202: ...2215A SERVICE ...
Page 205: ...000 2775 00D ...