![Tektronix 2215A Instruction Manual Download Page 68](http://html.mh-extra.com/html/tektronix/2215a/2215a_instruction-manual_1078831068.webp)
b.
Set the calibration generator to produce a 20-mV
signal.
c. Set the bottom of the signal on the center horizontal
graticule line using the Channel 2 POSITION control.
d. Set the Channel 2 Input Coupling switch to AC.
e.
CHECK— Display is centered about the center hori
zontal graticule line.
f.
Move the cable from the CH 2 OR Y input connector to
the CH 1 OR X input connector. Set the VERTICAL MODE
switch to CH 1.
g. Repeat parts c through e using the Channel 1 controls.
h. Disconnect the test equipment from the instrument.
9. Adjust Attenuator Com pensation (C12, C11, C5,
C4, C62, C61, C55, and C54)
a. Set:
VOLTS/DIV (both)
0.1 V
Input Coupling (both)
DC
A SEC/DIV
20
ns
b. Connect the high-amplitude square wave output via a
50-S2 cable and a 50-li terminator to the CH 1 OR X input
connector.
c.
Set the generator to produce a 10-kHz, 5-division
display.
NOTE
Use Table 5-4 to identify the correct capacitor for
each channel adjustment.
d.
ADJUST— The 10X LF Comp capacitor for best front
corner.
e. Set the CH 1 VOLTS/DIV switch to 1 V.
f. Set the generator to produce a 5-division display.
Adjustment Procedure— 2215A Service
Table 5-4
Attenuator Compensation Adjustments
Adjustment
Channel 1
Channel 2
10X LF Comp
C12
C62
10X Input C
C11
C61
100X LF Comp
C5
C55
100X Input C
C4
C54
g.
ADJUST— The 100X LF Comp capacitor for best front
corner.
h.
Replace the
50-Q
cable and 50-fi termination with the
10X probe (supplied with the instrument).
i.
Set the CH 1 VOLTS/DIV switch to 10 mV and the A
SEC/DIV switch to 0.2 ms.
j.
Insert the tip of the X I0 probe into the PROBE AD
JUST output jack.
k.
Adjust the X I0 probe compensation for a flat top on
the waveform. Refer to the instructions supplied with the
probe for details of compensation adjustment.
I.
Set the CH 1 VOLTS/DIV switch to 0.1 V or 1 V (1 OX
PROBE).
m.
Connect the X I0 probe tip to the high amplitude gen
erator via a probe-tip-to-BNC connector.
n. Set the generator to produce a 1 kHz, 5-division
display.
v
o. ADJUST— The 10X Input C capacitor for best flat top.
p.
Set the CH 1 VOLTS/DIV switch to 1 V or 10 V (X I0
PROBE) and set the generator to produce a 5-division
display.
q.
ADJUST— The 100X Input C Capacitor for the best
flat top.
r. Set the VERTICAL MODE switch to CH 2.
5-9
Summary of Contents for 2215A
Page 1: ......
Page 9: ...2215A Service 4735 01 vi The 2215A Oscilloscope ...
Page 138: ...2215A Service 4735 04 Figure 9 4 Basic block diagram ...
Page 139: ......
Page 145: ...CIRCUIT BCARD N TER CONNECTIONS 2 2 1 5 A 4755 20 ...
Page 153: ...2215A S e rv ic e D UJ o C D GO CO in 10 CJ Figure 9 8 Circuit view of A1 Main board ...
Page 157: ...1 p 2 m 3 p 4 p 5 P 6 p 7 p 8 p 9 A T B Z 2 I5 A ...
Page 158: ...H T J t K t L t M T N T p t S 4 7JS Z2 ...
Page 168: ...I 2 3 4 5 6 7 8 9 B H K L M N 2 2 I 5 A 473S Z ...
Page 173: ...I 2 to 3 to 4 to 5 to 6 7 to 8 to 9 2 215A B TIMING ALTERNATE B SWEEP S ...
Page 176: ......
Page 181: ...I 2 3 4 5 6 7 8 9 AXIS CRT ...
Page 189: ...2215A Service 4735 11 ...
Page 193: ...3 ...
Page 194: ...2215A SERVICE ...
Page 195: ...4 7 5 ...
Page 200: ......
Page 202: ...2215A SERVICE ...
Page 205: ...000 2775 00D ...