Section 4— 2215A Service
PERFORMANCE CHECK PROCEDURE
INTRODUCTION
PURPOSE
The “Performance Check Procedure" is used to verify the
Performance Requirement statements listed in Table 1-1. It
is the recommended acceptance check procedure for new
instruments.
Instrument performance should be checked after every
2000 hours of operation or once each year, if used infre
quently. A more frequent interval may be necessary, if your
instrument is subjected to harsh environments or severe us
age. The results of these periodic checks will determine the
need for readjustment.
Selected procedures may also be used as preliminary
troubleshooting aids or to verify instrument performance af
ter repair or component replacement.
STRUCTURE
This procedure is structured into four major subsections,
each of which can be performed independently, to permit
checking individual portions of the instrument. At the begin
ning of each subsection there is an equipment-required list
showing only the test equipment necessary for performing
the steps in that subsection. In this list, the Item number
that follows each piece of equipment corresponds to the
Item number listed in Table 4-1.
Also at the beginning of each subsection is a list of all the
front-panel control settings required to prepare the instru
ment for performing Step 1 in that subsection. Each suc
ceeding step within a particular subsection should then be
performed, both in the sequence presented and in its en
tirety, to ensure that control-setting changes will be correct
for ensuing steps.
TEST EQUIPMENT
The test equipment listed in Table 4-1 is a complete list of
the equipment required to accomplish both the “Perfor
mance Check Procedure" in this section and the ‘ Adjust
ment Procedure" in Section 5. To assure accurate
measurments, it is important that test equipment used for
making these checks meet or exceed the specifications de
scribed in Table 4-1. When considering use of equipment
other than that recommended, utilize the “Minimum Speci
fication" column to determine whether available test equip
ment will suffice.
Each procedure in this section is written using the control
and connector nomenclature imprinted on the “recom
mended" test equipment. When substitute equipment is
used, control settings stated in the test setup and in the
procedure itself may need to be altered.
Detailed operating instructions for test equipment are not
given in this procedure. If more operating information is re
quired, refer to the appropriate test-equipment instruction
manual.
LIMITS AND TOLERANCES
The tolerances given in this procedure are valid for an
instrument that is operating in and has been previously cali
brated in an ambient temperature between + 20°C and
+ 30°C. The instrument also must have had at least a 20-
minute warm-up period. Refer to Table 1-1 for tolerances
applicable to an instrument that is operating outside this
temperature range. All tolerances specified are for the in
strument only and do not include test-equipment error.
4-1
Summary of Contents for 2215A
Page 1: ......
Page 9: ...2215A Service 4735 01 vi The 2215A Oscilloscope ...
Page 138: ...2215A Service 4735 04 Figure 9 4 Basic block diagram ...
Page 139: ......
Page 145: ...CIRCUIT BCARD N TER CONNECTIONS 2 2 1 5 A 4755 20 ...
Page 153: ...2215A S e rv ic e D UJ o C D GO CO in 10 CJ Figure 9 8 Circuit view of A1 Main board ...
Page 157: ...1 p 2 m 3 p 4 p 5 P 6 p 7 p 8 p 9 A T B Z 2 I5 A ...
Page 158: ...H T J t K t L t M T N T p t S 4 7JS Z2 ...
Page 168: ...I 2 3 4 5 6 7 8 9 B H K L M N 2 2 I 5 A 473S Z ...
Page 173: ...I 2 to 3 to 4 to 5 to 6 7 to 8 to 9 2 215A B TIMING ALTERNATE B SWEEP S ...
Page 176: ......
Page 181: ...I 2 3 4 5 6 7 8 9 AXIS CRT ...
Page 189: ...2215A Service 4735 11 ...
Page 193: ...3 ...
Page 194: ...2215A SERVICE ...
Page 195: ...4 7 5 ...
Page 200: ......
Page 202: ...2215A SERVICE ...
Page 205: ...000 2775 00D ...