Performance Check Procedure—2215A Service
c. Press in the SGL SWP RESET button. The READY
LED should illuminate and remain on.
f. Press in the SGL SWP RESET button several times.
d. Set the Channel 1 Input Coupling switch to DC.
e. CHECK— READY LED goes out and a single sweep
occurs.
g. CHECK— Single-sweep trace occurs, and the READY
LED illuminates briefly every time the SGL SWP RESET
button is pressed in and released.
NOTE
The A IN TEN SITY con tro l m ay require adjustm ent to
observe the single-sw eep trace.
h. Disconnect the test equipment from the instrument.
4-13