Adjustment Procedure— 2215A Service
2. Check High-Voltage Supply
W A R M I N G
Instrum ent m u st be turned o ff when rem oving o r re
placing the c rt cover and cap.
a. Remove the crt cover and cap and connect a dc volt
meter capable of measuring at least —2500 V between pin
2 of the crt socket and chassis ground. Pin 2 of the crt is
negative with respect to the chassis.
b. CHECK— Voltmeter reading is between —1900 V and
-2 1 0 0 V.
c. Disconnect the voltmeter leads and replace the crt cap
and cover.
3. Adjust CRT Grid Bias (R851)
a. Connect a 50-Q termination to the EXT Z AXIS INPUT
connector located on the rear panel.
b. Adjust the front-panel FOCUS control to produce a
well-defined dot.
c. Rotate the A INTENSITY control fully counterclock
wise.
d. ADJUST— Grid Bias (R851) for a visible dot. Then
back off the Grid Bias potentiometer until the dot just
disappears.
e. Disconnect the 50-Q termination from the EXT Z AXIS
INPUT connector.
4. Adjust Astigm atism (R874)
a. Set:
A INTENSITY
Visible display
Channel 1 Input Coupling
DC
A SEC/DIV
5
ms
b. Connect the leveled sine-wave generator output via a
50-Q cable and a 50-Q termination to the CH 1 OR X input
connector.
c. Set the generator to produce a 50-kHz, 4-division
display.
d. ADJUST— Astig (R874) and the front-panel FOCUS
control for the best defined waveform.
e. Disconnect the test equipment from the instrument.
5. Adjust Trace Alignment
a. Set the Channel 1 Input Coupling switch to GND and
position the trace to the center horizontal graticule line.
b. ADJUST— The front-panel TRACE ROTATION con
trol for optimum alignment of the trace with the center hori
zontal graticule line.
6. Adjust Geom etry (R870)
a. Set:
CH 1 VOLTS/DIV
50 mV
CH 1 Input Coupling
DC
A SEC/DIV
0.1 ms
b. Connect
2 0 -
m s
time markers from the time-mark gen
erator via a 50-Q cable and a 50-Q termination to the CH 1
OR X input connector.
c. Adjust the Channel 1 POSITION control fully counter
clockwise to position the baseline part of the display below
the bottom horizontal graticule line.
d. ADJUST— Geom (R870) for minimum curvature of the
time markers at the left and right edges of the graticule.
e. Set the Channel 1 Input Coupling switch to GND.
f. ADJUST— Geom (R870) for minimum curvature of the
baseline trace when positioned at the top and bottom hori
zontal graticule lines using the Channel 1 POSITION
control.
g. Set the Channel 1 Input Coupling switch to DC.
h. Repeat parts d through g for optimum compromise
between the vertical and horizontal displays.
i. Disconnect the test equipment from the instrument.
5 - 5
Summary of Contents for 2215A
Page 1: ......
Page 9: ...2215A Service 4735 01 vi The 2215A Oscilloscope ...
Page 138: ...2215A Service 4735 04 Figure 9 4 Basic block diagram ...
Page 139: ......
Page 145: ...CIRCUIT BCARD N TER CONNECTIONS 2 2 1 5 A 4755 20 ...
Page 153: ...2215A S e rv ic e D UJ o C D GO CO in 10 CJ Figure 9 8 Circuit view of A1 Main board ...
Page 157: ...1 p 2 m 3 p 4 p 5 P 6 p 7 p 8 p 9 A T B Z 2 I5 A ...
Page 158: ...H T J t K t L t M T N T p t S 4 7JS Z2 ...
Page 168: ...I 2 3 4 5 6 7 8 9 B H K L M N 2 2 I 5 A 473S Z ...
Page 173: ...I 2 to 3 to 4 to 5 to 6 7 to 8 to 9 2 215A B TIMING ALTERNATE B SWEEP S ...
Page 176: ......
Page 181: ...I 2 3 4 5 6 7 8 9 AXIS CRT ...
Page 189: ...2215A Service 4735 11 ...
Page 193: ...3 ...
Page 194: ...2215A SERVICE ...
Page 195: ...4 7 5 ...
Page 200: ......
Page 202: ...2215A SERVICE ...
Page 205: ...000 2775 00D ...