Triggering on Waveforms
TDS 500D, TDS 600B, & TDS 700D User Manual
3–117
Level lets you enter the trigger level using the general purpose knob or the
keypad.
Set to TTL fixes the trigger level at +1.4 V.
Set to ECL fixes the trigger level at –1.3 V.
Set to 50% fixes the trigger level to approximately 50% of the peak-to-peak
value of the trigger source signal. When AMI is selected, this selection measures
the peak-to-peak level and sets the upper threshold to 75% and the lower
threshold to 25%. If you select a DS0 mask the trigger level is set correctly, do
not press the front panel button SET LEVEL TO 50%.
Delayed Triggering
The TDS Oscilloscope provides a main time base and a delayed time base. The
delayed time base, like the main time base, requires a trigger signal and an input
source dedicated to that signal. You can only use delay with respect to the main
edge trigger and certain classes of main pulse triggers. This section describes
how to delay the acquisition of waveforms.
There are two different ways to delay the acquisition of waveforms: delayed runs
after main and delayed triggerable. Only delayed triggerable uses the delayed
trigger system. Delayed runs after main looks for a main trigger, then waits a
user-defined time, and then starts acquiring. (See Figure 3–54.)
Wait User-specified
Time
Acquire
Data
Wait for
Main
Trigger
Figure 3–54: Delayed Runs After Main
Delayed triggerable looks for a main trigger and then, depending on the type of
delayed trigger selected, makes one of the three types of delayed triggerable mode
acquisitions: After Time, After Events, or After Events/Time. Study Figure 3–55 to
understand the sequence the oscilloscope goes through for each delayed mode.
Summary of Contents for 071-0130-00
Page 4: ......
Page 12: ...Table of Contents viii TDS 500D TDS 600B TDS 700D User Manual ...
Page 18: ...Preface xiv TDS 500D TDS 600B TDS 700D User Manual ...
Page 19: ...Getting Started ...
Page 28: ...Operating Basics ...
Page 30: ...Overview 2 2 TDS 500D TDS 600B TDS 700D User Manual ...
Page 60: ...Tutorial 2 32 TDS 500D TDS 600B TDS 700D User Manual ...
Page 61: ...Reference ...
Page 65: ...Overview 3 4 TDS 500D TDS 600B TDS 700D User Manual ...
Page 185: ...Triggering on Waveforms 3 124 TDS 500D TDS 600B TDS 700D User Manual ...
Page 257: ...Saving Waveforms and Setups 3 196 TDS 500D TDS 600B TDS 700D User Manual ...
Page 300: ...Appendices ...
Page 310: ...Appendix A Options and Accessories A 10 TDS 500D TDS 600B TDS 700D User Manual ...
Page 328: ...Appendix B Algorithms B 18 TDS 500D TDS 600B TDS 700D User Manual ...
Page 330: ...Appendix C Packaging for Shipment C 2 TDS 500D TDS 600B TDS 700D User Manual ...
Page 340: ...Appendix F Programmer Disk F 2 TDS 500D TDS 600B TDS 700D User Manual ...
Page 341: ...Glossary ...
Page 355: ...Glossary Glossary 14 TDS 500D TDS 600B TDS 700D User Manual ...
Page 356: ...Index ...
Page 374: ...Index Index 18 TDS 500D TDS 600B TDS 700D User Manual ...
Page 375: ......
Page 376: ......