Technology Solutions
Instruction Manual
Tek-Wave 4300C
www.tek-trol.com
29
5.3.3.6
Copy Sensor Data
•
In this menu you can back up the sensor settings to restore them when necessary.
•
Copy from sensor means to save the sensor settings and copy to sensor means to restore the
sensor settings.
5.3.3.7
PIN
•
In this menu, the PIN is activated or inactivated permanently.
•
Entering a 4-digit PIN to protect the sensor data against unauthorized access and
unintentional modifications.
5.3.3.8
Distance Adjustments
•
Distance adjustments is used to correct the difference between the measured value and
•
actual distance.
•
Press to enter Distance Adjustments.
5.3.3.8.1
Threshold Setting
•
Threshold setting is used to set the significant size of the echo threshold.
•
The threshold set higher to respond effectively to fluctuations in Degree, which excludes the
small-signal clutter.
*
Note: If you modify the threshold value is greater than the effective echo amplitude, the wave will cause
misunderstanding. The menu includes the echo threshold and amplitude envelope, which is the default rate of return
threshold for the 60mV, amplitude envelope default value of 10mV.