ACA:
Range
Resolution
Accuracy
600
A
0.1
A
±(1.8% of rdg+5 digits)
6000
A
1
A
60mA
0.01mA
±(1.8% of rdg+5 digits)
600mA
0.1mA
6A
1mA
±(3.0% of rdg+8 digits)
10A
10mA
Above accuracies can be guaranteed within 5%~100%
of the full range.
The true RMS meter has residual value within 10 counts
when the test leads are shorten, but that will not affect
the accuracy of measurement.
1. Frequency Range for ACA: 40Hz-400Hz
2. Overload protection: F 10A/1000V fuse for 10A
Overload protection: F 0.63A/1000V fuse for
A and mA
ranges.
3. Maximum input current: 600mA dc or 600mA ac rms for
A
and mA ranges, 10A dc or 10A ac rms for 10A ranges.
4. For measurements>6A, 15 seconds ON each 10 minutes;
Above 10A unspecified.
5. MAINTENANCE
This section provides basic maintenance information,
including fuse and battery replacement instructions.
Do not attempt to repair or service your Meter unless you are